Loose communication

Check adjustment

Large vibration after startup

The centrifugal pipe in the rotor is

not placed symmetrically

Check adjustment

Rupture of centrifugal pipe

Inspection and

replacement

Rotor not tightened

Check adjustment

Shock absorption part damaged

Check and replace

The display shows 0000, press the

start key and the machine will not

run

Circuit board or transformer damaged Replace

Loose connector of control system

Reinsert

The key is damaged

Replace panel

Motor damage or leakage

Replace

It can run but can't go up in speed.

The machine has strange noise or

peculiar smell

Control system or motor failure

Send to the

manufacturer for

maintenance

The actual speed is different from

the set or displayed speed, out of

control or unstable

Control system failure

Send to the

manufacturer for

maintenance

The temperature does not meet the

set low temperature requirements

Without precooling or insufficient

Precooling

Sufficient

Precooling

Compressor damaged

Check and replace

amaged solenoid valve

Check and replace

Insufficient refrigerant

Add proper

amount of

refrigerant

Summary of Contents for HZDD5Y

Page 1: ...HZDD5Y Centrifuge Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...n technical parameters 2 IV Technical parameters of rotor 3 V Shape and mechanism 3 VI Installation 3 VII Operation method 4 VIII Precautions 7 IX Maintenance 8 X Common faults and troubleshooting 8 X...

Page 3: ...p automatically and give an alarm to ensure the safety of the instrument in use The instrument adopts the optimized design scheme for the shock absorption system The instrument operates stably with sm...

Page 4: ...n farthest from the axis Rmin radius of rotation of the test solution nearest to the axis density of mixed liquid g cm3 viscosity of mixture poise N speed RPM R particle radius CM particle density g c...

Page 5: ...hape and mechanism As shown in the figure VI Installation After the instrument arrives at your company check whether the appearance of the packing box is severely damaged in the process of transportat...

Page 6: ...heating pipe and radiator etc 2 Installation space requirements The centrifuge should have enough space to ensure the air flow around the centrifuge otherwise it may lead to overheating and poor effe...

Page 7: ...inspection if conditions permit balance weight is recommended 2 If the reagent is not enough empty test tube shall be used for water injection instead of unbalanced operation 3 Finish the work above...

Page 8: ...e window flashes the time setting will be entered and then press the plus or minus key to determine the running time 4 When the above steps are completed press the approval key in time to confirm othe...

Page 9: ...l sound before leaving 5 During the maintenance of the instrument the power plug shall be removed otherwise electric shock may occur easily Tips 1 The hanging cup shall be placed on the rotor hanging...

Page 10: ...om the centrifugal chamber apply a thin layer of grease to the center hole of the rotor and the main shaft of the centrifuge and keep the rotor in a dry place X Common faults and troubleshooting Some...

Page 11: ...t The key is damaged Replace panel Motor damage or leakage Replace It can run but can t go up in speed The machine has strange noise or peculiar smell Control system or motor failure Send to the manuf...

Page 12: ...is mainly used to protect the semiconductor elements of the instrument control system The fuse is a closed glass tube structure and is connected by plug in type It has compact structure safe use and...

Page 13: ...llowing situations of supporting rotors Use when the maximum speed of rotor is exceeded The rotor is scratched or bruised The rotor is seriously corroded by acid and alkali Repair by unauthorized prof...

Reviews: