background image

6

3.1.2 Voltage adjustment

Turn the voltage adjusting wheel clockwise to the end, the pointer will be from 0 to maximum,

otherwise the pointer will be from maximum to 0.

3.1.3 Choose stirring rate

Turn on “RUN”, stir motor starts working. Observe if the stirrer blade is flexible and if stirring is even.

Turn stir adjusting wheel to adjust stirring rate from 0-40 roll/min, for example, if you need adjust to

30 roll/min, you can rotate to 7.5 scale point.

3.1.4 Igniting

Push down “IGNITION” button on control panel, shutter slides and door opens, igniter descends to

test cup’s vapor space. When igniting, stir motor automatically stops.

3.1.5 Input medium

If flash point of sample around 15

, user can use icy water as medium, and if flash point around

40

, water with normal temperature can be used. Switch off the water inlet cover and funnel the

medium into the bath. Slow the speed in the late of stage and observe if any liquid flow from liquid

return pipe. Pay attention, if too fast, medium might flow out from thermometer hole or water inlet

hole.

If flash point of sample at range of -30

to 15

, additional cooling bath should be used

Connect intake pipe to In/Out liquid joint, connect liquid return pipe to overflow pipe and fix well.

Open the stop valve slightly, and then open cooling bath, input medium slowly and observe if any

liquid flow out. If observed, shut cooling bath and valve immediately.

Note: Circulating rate must not too fast, otherwise cold liquid might flow out and hurt operator.

Water inlet hole
with cover

Summary of Contents for HZBS-002

Page 1: ...HZBS 002 Abel Flash Point Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...nd operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrume...

Page 3: ...Contents I General 1 II Installation and commissioning 2 III Operation 5 VI Trouble and trouble shooting 7 V Maintenance 8...

Page 4: ...d by SSR which can satisfy requirements of different heating rate Continuously stirring the sample and stirring is auto stopped during igniting Electrical controlled ignition Test cup lid with stirrin...

Page 5: ...mperature rise rate Stirring rate can be adjusted in range of 0 40 roll min by stirrer adjust button Stirrer will be stop working during ignition 1 3 3 Gas source Beside master gas valve we have adjus...

Page 6: ...2 2 1 The instrument should be placed on steady test platform with no airflow around 2 2 2 The distance between instrument and the walls or other shields should be not less than 200mm 2 3 Installation...

Page 7: ...t gas source into gas inlet and fix well If test in low temperature connect intake pipe to In Out liquid joint connect liquid return pipe to overflow pipe and fix well If test not in low temperature y...

Page 8: ...alve and adjusting valves connect the gas source with the master gas valve When igniting turn on the master valve Fully or half open depending on the gas pressure and turn left the igniting adjusting...

Page 9: ...point of sample around 15 user can use icy water as medium and if flash point around 40 water with normal temperature can be used Switch off the water inlet cover and funnel the medium into the bath S...

Page 10: ...will be descended 3 2 4 When test operation is complete adjust voltmeter back to 0 turn off RUN switch off gas source remove all liquid medium in the bath VI Trouble and trouble shooting Trouble Caus...

Page 11: ...haft and stationary shaft homocentric Fire go out when ignition 1 Test flame too small 2 Disturbing air flow 1 Adjust flame size 2 Prevent air flow V Maintenance 5 1 The instrument should be placed at...

Reviews: