Use instrument charger to recharge
USB interface
External USB disk is used to store test data, please use FAT or
FAT32 format U disk; in the storage process, it is strictly prohibited to
dial out the USB disk.
IX.Operation Instructions
Test wiring
Two test clamps are clamped to switch contacts or test ends.
I+/U+terminals are connected with red test lines, I-/U-terminals are
connected with black test lines, in which the thick-wire inserts are
connected with current terminal I+/I-, and the thin-wire rods are inserted into
voltage terminal U+/U-.
It is necessary to use the special low resistance test line matched with the
instrument, tighten the current connection pole, and firmly clamp the test
clamp to reduce the lead resistance and save the battery power.
If the test line contacts badly and the current cannot reach the set value, the
instrument will stop testing and alarm.
Test grounding
In live environment, it is necessary to ensure that one end of the sample
has been grounded.
Intelligent Power Management
When the instrument is not operated for a long time, it shuts down
automatically to save electricity; the instrument has low charge prompt
function and over-discharge protection function.
Instructions for use of printer
The key of the printer and the indicator of the printer are integrated. When
the printer is powered on, the indicator light is normally on, and it flashes
when the paper is missing. Press the button once and the printer passes
the paper.