benzoic acids is calculated and compared with the standard calorific value, and the error
is less than 120J, which means that the calibration is successful.
(7) Calorific value measurement
Weigh the sample (0.9 ~ 1.1g), load the oxygen bomb, charge oxygen for 15 ~ 45
seconds, pressure is 2.8 ~ 3.2Mpa, put it into the inner cylinder, click "calorific value test",
the instrument will start calorific value measurement, input the weight of weighed sample,
input the weight of additive, if there is no additive, you can input 0. The results are
displayed and printed after the test.

VII.Common faults and troubleshooting

phenomenon

reason

handle

1

Oxygen bomb

leak

Aging or wear of rubber sealing ring Replace the sealing ring

2

Ignition failure

The circuit is blocked or the

ignition wire is in poor contact

The sample is wet

Poor contact between ignition

wire or cotton thread and sample

The two electrodes are too dirty

Oxidation of ignition cap

The two electrodes are short

circuited with the crucible (it is easy
to burn the crucible and electrode at
this time)

Check whether the wiring is well

connected, whether the oxygen
warhead is in good contact with the
ignition cap, and whether the oxygen
bomb is placed properly

The sample is splashed with

oxygen too fast

Re sample

Sand the electrode with

sandpaper

Sand the ignition cap oxide with

sandpaper

Replace the electrode or crucible

and re load the sample

3Incomplete
combustion of
sample

The sample is not flammable
Insufficient oxygen or insufficient
oxygen pressure

Wrap the sample with lens wiping
paper, extend the oxygen filling time,
and replace the oxygen cylinder

4

After ignition,

the temperature
rises too high and
the calorific value
is too high

Agitator does not work

Mixing blade falling off

The mixing shaft is stuck and the

line is blocked

Insert a cotton into the nylon rod

connected with the mixing shaft and
insert it again

5

The test did not

end for a long time

The ambient temperature is too
high

Adjust the water temperature of the
outer cylinder to the room
temperature, or lower the indoor
temperature

6

Air leakage of

oxygenator

Aging or wear of sealing ring of
oxygenator

Replace the sealing ring

Summary of Contents for HZ-384A

Page 1: ...HZ 384A Automatic Calorific Value Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...al and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this inst...

Page 3: ...nts I Overview 1 II Technical features 1 III Working principle 1 IV Technical index 1 V Instrument structure and auxiliary equipment 2 VI How to use the instrument 2 VII Common faults and troubleshoot...

Page 4: ...emperature recovery correction calculation and printing are completed automatically to avoid human error and improve the accuracy and precision 3 3The unique calibration technology makes the requireme...

Page 5: ...ser interface and so on 2 The water tank is made of stainless steel and plastic VI How to use the instrument 1 Preparation before use Open box inspection after unpacking check whether the appearance o...

Page 6: ...strument 2 Instructions for use of the instrument The main interface of the instrument is as follows Figure 1 1 system system click system setting to enter the system setting menu as shown in the figu...

Page 7: ...of the instrument is directly input according to the data on the package of the reference material used Click return to save and return to the main interface of the system water injection time click...

Page 8: ...matic printing that is the test results will be printed automatically after the test 1 It means manual printing The result will not be printed after the test but can be printed by clicking Print manua...

Page 9: ...on the ignition circuit To detect the ignition circuit Click the mixing test and the MCU sends out the mixing command to switch on the mixing It is used to detect the stirring circuit Click water inj...

Page 10: ...o the following Click calorific value input as follows Input the weight of the sample if no additives are used you can input 0 serial number hydrogen sulfide water and other experimental data and clic...

Page 11: ...e processing trace so as not to affect the print results 5 reset Press this key and the instrument will return to the power on state 6 Thermal capacity calibration Weigh benzoic acid 0 9 1 1g load oxy...

Page 12: ...cible it is easy to burn the crucible and electrode at this time Check whether the wiring is well connected whether the oxygen warhead is in good contact with the ignition cap and whether the oxygen b...

Page 13: ...ng instrument 1 5 Pressure gauge 1 6 Ignition wire 3 7 Cotton 1 8 Benzoic acid 1 9 Add drain pipe 1 10 Tweezers 1 11 Printing paper 2 12 Oxygen bomb support 1 13 crucible 5 14 Oxygen bomb sealing ring...

Reviews: