Grounding post: the grounding point of the whole instrument for safety protection
I+, I-: Output current binding posts, I+ is the output current is positive, I- is the output
current is negative.
V+, V-: Voltage sampling terminal, V+ is the positive terminal of the voltage line, and V-
is the negative terminal of the voltage line.
RS232: Universal serial interface, the instrument can be controlled by a computer.
USB: Test results can be output to U disk.
Charging socket: AC 220V power supply, the battery can be charged after plugging in
the charger
Charging indication: The green indicator light indicates the charging is complete, and
the red indicator indicates the charging process.

V.Test and operation method

Take out and connect the special test leads provided with the machine. The red and
black test pliers are respectively clamped to two of the down conductors of the
grounding grid, and rub the contact points hard to ensure good contact. The other end
of the test line Correspond to the wiring terminal of the instrument. After confirming that
the test line is connected correctly, connect the power line and prepare for
measurement. At this time, turn on the power switch and the following interface is
displayed:
1. The boot interface is shown in Figure 1:

Figure 1

Summary of Contents for HZ-3510

Page 1: ...HZ 3510 Grounding Continuity Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this instrum...

Page 3: ...nd characteristics of the instrument 1 II Main Specifications 1 III Main features 2 IV System introduction 2 V Test and operation method 3 VI Four terminal wiring method 6 VII Precautions 6 VIII After...

Page 4: ...Prevent Major Accidents in Electric Power Production it is clearly stated that the conduction test of the down conductor of the grounding device should be carried out once a year The HZ 3510 groundin...

Page 5: ...d new power supply technology with many current gears and wide measuring range 3 Intelligent power management technology the instrument always works in the minimum power state effectively reducing the...

Page 6: ...harger Charging indication The green indicator light indicates the charging is complete and the red indicator indicates the charging process V Test and operation method Take out and connect the specia...

Page 7: ...e test it will display Charging please wait to enter the test state a few seconds later the test result will be displayed as shown in Figure 3 Figure 3 After the test result is displayed click local s...

Page 8: ...cursor to the clock modification press the select key to enter the clock modification interface As shown in Figure 5 Figure 5 Click to set the date and time 5 Click System Settings in Figure 1 to ent...

Page 9: ...ur terminal wiring method As shown in Figure 7 Figure 7 VII Precautions 1 When selecting the current please refer to the range in the technical index column When the range is exceeded the instrument i...

Page 10: ...aired free of charge for product quality problems belonging to the company within one year from the date of purchase and lifetime warranty and technical services will be provided If you find that the...

Reviews: