light is off

Poor power connection

Check the circuit for short

circuit

The power switch is damaged

Replace the power switch

Turn on the

water bath

stirring switch,

the motor does

not rotate

There is a problem with the

stirring switch

Replace the stirring switch

There is a problem with the

stirring motor

Replace the mixing motor

Open circuit

Check and exclude

The bath

temperature is

not constant

Solid state relay is broken

Replace solid state relay

Damaged heating tube

Replace heating tube

Sensor damaged

Replace the sensor

Temperature control table is

damaged

Replace the temperature

control table

The timer does

not show

Timer open

Check the line

The timer does

not count or

clear

Bad timer contact

Check the connection

Timer malfunction or button

switch is broken

Replace timer or button

switch

VI. Precautions

1. Keep the instrument clean, do not get water or damp into the electronic control part,

and prevent dirt such as oil samples from being spilled on the instrument.

2. The instrument should be well grounded before use to ensure the safety of operators.

3. When the instrument fails, please professional and technical personnel to deal with it,

and do not disassemble it at will.

4. This instrument is guaranteed for one year.