21
Table9- 1 The Calibrating Points for Sourcing Functions
Source
Calibrating Point
Zero
Span
VDC
0V
10V
——
——
——
——
——
——
——
——
mVDC
0mV
100m
V
——
——
——
——
——
——
——
——
mADC
1mA 20mA
——
——
——
——
——
——
——
——
Ω
15Ω
40Ω
40Ω
400Ω 400Ω 800Ω
800Ω
1800
Ω
1800Ω
3200Ω
Table9- 2 The Calibrating Points for Measurement Functions
Measure
Zero
Span
Source/Measure VDC
1V
20V
Source/Measure mVDC
1mV
100mV
Source/Measure Ω
0Ω
400Ω
500Ω
3000Ω
Measure VDC
0.1V
20V
Measure mVDC
1mV
100mV
Measure mADC
0.1mA
20mA
VIII.Clear Function
Before sourcing or measuring DC voltage, resistance, RTDS and TC, offset errors
caused by internal elements or external factors must be cleared.
CAUTION: The cleared values will be invalid at power on, except sourcing resistance and RTDs.
CAUTION: The cleared range as follows:
V (-10mV~+10mV),
mV and TC (-50μV~+50μV),
Resistance and RTD (-10Ω~+10Ω).
8.1 Clear of Measuring Voltage
Clear of Measuring V DC in upper display:
Press
until the upper display shows “V”. Short the test leads with Measure jacks
and press left
.
Clear of Measuring mV DC in upper display:
Press
until the upper display shows “mV”. Short the test leads with Measure jacks
and press left
.
Clear of Measuring V DC in lower display:
Press
until the upper display shows “V”. Short the test leads with Source/Measure