Setting of system parameters:
Press “Enter” key and enter the following interface:
Setting of voltage boosting: the user may select based on actual demand.
Voltage of boosting stop: 10
~
100KV
Autotype: Yes/ No
Standing time: 0
~
15min
Interval of boosting: 0
~
5min
System time:
Stirring switch: On / Off
Boosting frequency: 1
~
6
Voltage boosting rate: 2.0KV/S
~
3.5 KV/S
Press “Esc” key to exit this interface after the setting is done.
③
Testing:
Press the “Select” key to select the “Start Test” menu and press the “Enter” key to enter
the following interface: