background image

6

Press Cycle key the cursor can move among

Current select

,

Phase division

,

Winding

temp

,

Convert temp

,

Data query

,

Clock modify

,

Parameters

,etc. Press select key

can circle select above 7 menus option. when the cursor at Winding temp ,press start key

the cursor will roll between 3 data bits ,the

select

key can circle display data between

0~9.the Start up key can start any options besides Winding temp.

3. Move the cursor to data query menu at the boot state, then press select key enter to the

data query ,press select or start up key to turn pages querying.

4. After selecting the current, press START UP to start the test; When press START UP,

few seconds later, it will display "CHARGING", when it displays "TESTING" that means

the “CHARGING” has been completed. Few seconds after accessing the TEST STATUS it

will show the measured resistance values.

5. After the test: press "RESET" back to boot interface, discharge, alarm,and after the

discharge alarm, please be sure wait for about 10 seconds, re-connection, proceed the

next test or unplug the test cable and end the measurement.

Summary of Contents for HZ-3140A

Page 1: ...HZ 3140A Transformer DC Resistance Tester Huazheng Electric Manufacturing Baoding Co Ltd...

Page 2: ...ual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this ins...

Page 3: ...s I Introduction 1 II Safety Measure 1 III Configuration 1 IV Technical Data 2 V Presentation Of The System 3 VI Test And Operating 3 VII Caution 7 VIII Guarantee and After Sales Service 7 IX Packing...

Page 4: ...s It is a test instrument of high precision simple direct way to achieve fast transformer measurement II Safety Measure 1 Before using this equipment you must carefully read the manual 2 The equipment...

Page 5: ...agement technology instrument is always working at minimum power state effective energy saving reduce heat 7 320 x240 matrix of ultra small pixel 65K true color LCD 8 With calendar clock and power sto...

Page 6: ...selecting the current The microcomputer control to realize all testing process I I Current output terminals I is the positive current I is the negative current V V Voltage input terminals V is the pos...

Page 7: ...A B C I I Testing product Measurement device O V V a b c B magnetic assisted connections see picture 3 to 5 suitable Y N d 11junction group A B C I I Measurement device O V V a b c Testing product Pic...

Page 8: ...device O V V a b c Testing product A B C I I Measurement device O V V a b c Testing product Picture 3 4 5 respectively measure low voltage Rac Rba Rcb connection 2 Instrument boot up page is as follow...

Page 9: ...menu at the boot state then press select key enter to the data query press select or start up key to turn pages querying 4 After selecting the current press START UP to start the test When press START...

Page 10: ...ll display CURRENT VERY LOW when the two status appear to confirm the range and select the suitable current and carry testing 4 After the test press RESET and winding device will disconnect the power...

Page 11: ...IX Packing List No Item Qty 1 Main engine 1 2 Red test line 1 3 Black test line 1 4 Power line 1 5 Ground lead 1 6 Fuse pipe 2 7 Print paper 1 8 Test sample Standard resistance 1 9 Serial port line 1...