OUTPUT jack shall not be connected with
the voltage higher than 30V to avoid
damage to the instrument.
4.21.1 Turn the rotary switch until it leaves OFF to power on the
instrument.
4.21.2 Press SHIFT to make all keys under used for output state.
4.21.3 Press FUNC to select output signal type (such as voltage
output).Then the output display area of LCD shows type ―D
C‖ with measuring unit ―V‖.
4.21.4 The voltage output has two scales: 0
~
10V and 0
~
100mV, which
can be switched using RANGE key to ensure different
resolutions.
4.21.5 Respectively plug the red test clip and the black test clip into
and OUTPUT—. The former is the positive terminal
of the voltage while the latter is the negative terminal.
4.21.6 The magnitude of signal shall be adjusted with key ―
▲
‖, ―
▼
‖, ―
▲
‖
and ―
▼
‖.
4.21.7 The reading showed in the measurement display area of LCD is
the magnitude of current output voltage.
H782 ProcessMultimeter
Operation guidance
Note:
1) There is overload protection for the output, the output voltage load
shall not be higher than 20mA.
2) When adjusting the signal, click ―
▲
‖ or ―
▼
to change one or two
figures of the output signal approximatively. When pressing and