Installation and Instruction
2
2.2.2 - Setting Parameters Description
Setting Parameter
Text
Description
Name
English letters or numbers
Name of the logger
SN
10 characters
Must be the same as the one in the
label of logger
Logging Interval (secs)
sampling Interval (secs)
Number from 2 to 86400
Number from 1 to 240
Logging frequency
Sampling frequency in LCD screen
Offset
Channel Name
Input positive number to decrease the
value.input negative number to increase
the value.
Name of each channel
Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications. However, there may be times when you wish to
adjust the calibration of your logger. Logpro provides you with the ability to perform a single point offset calibration. This calibration can be used to
increase the accuracy of the logger for a restricted data range.
Name
S220-T8
HS220CW001
Data Logger
S220-T8
H 1103
10
0
86000
0.0
0.0
SN
Type
Model
Firmware
Logging Interval
(
Secs
)
60
Sampling Interval (Secs)
Logs
Total
CH1
CH2
0.0
CH3
0.0
CH4
0.0
CH5
0.0
CH6
0.0
CH7
0.0
CH8
A.System
B.Sampling
C.Storage
D.Offset
1
2
Name1
Name2
3
Name3
4
Name4
5
Name5
6
Name6
7
Name7
8
Name8
E.Channel Name
English Letters or Numbers
Total
Logs
Number 86000
Number from 0 to 86000
The total capacity of the logger’s storage
The count of records in the memory
Calibration mode: For example, pls fill
in 1.5 if the displayed temperature is
higher than the actual temperature of
1.5
℃
; fill in -1.5 if the displayed
temperature is lower than the actual
temperature of 1.5
℃
. If you need to
adjust the humidity, do the same, the
number field is -100~100.