Attention
3
• Plastic shell is ABS, should be retardant, not resistant to acid and
alkali.
• If repair is needed, only authorized technician can do the repair.
• The instrument configuration by 9V block battery is not rechargeable.
• USB power supply (4.8V) cannot let the Data Logger work.
Section 4.1
-
LCD Screen Dim
Insufficient battery or the environment temperature is too low or too high.
Solution
: In the case of insufficient battery, please replace the battery. If results
from environment temperature, please immediately take the logger out of the
environment.
Section 4.2
-
Data & Time Error
Reason:
• The Battery level is low.
• The Data Logger is not synchronous properties, before start recording.
Solution:
• Replace the battery.
• Please sync the properties, before start recording.
Section 4.3
-
Software “Runtime Error”
Reason:
• OS forbid software creating files.
Solution:
• Run the program (software) as an administrator.
• Install software in D or E drive if available.
• A data logger name cannot contain any of the following characters: \ / :
* ? “ < > |
• Software’s installation path cannot include Chinese character or garbage
character.
Section 4.4
-
Check COM Port Number
Press “Win” + “R” in keyboard -> Run “devmgmt.msc” to Open “Device
Manager” in Windows-> Expand “Ports (COM & LPT)” -> “USB-serial CH340
(COM No.)” is the Data Logger.