Input impedance: at 400mV range >40M

, at other ranges are 

10M

Overload protection: 1000V DC or 750V AC peak value. 

3.2.4 AC voltage (ACV) 

Range Accuracy 

Resolution 

400mV 

±(1.5% 5

0.1mV 

4V 

±(1.0% 5

1mV 

40V 10mV 

400V 100mV 

600V 1V 

Input impedance: at 400mV range >40M

, at other ranges are 

10M

Overload protection: 1000V DC or 750V AC peak value. 

Frequency response: at 600V range: 40~100Hz,   

at other ranges:40~400Hz. 

Display: average value response (based on sine wave RMS). 

3.2.5 AC current (ACA) 

Range Accuracy 

Resolution 

400A <40A: 

±(3.0% 

5

   

>40A: ±(2.0% 

5

 

100mA 

600A 1A 

Overload protection: >800A, input time<1 minute. 

Frequency response: 40Hz-200Hz. 

Display: average value response (based on sine wave RMS). 

3.2.6 Resistance 

Range Accuracy 

Resolution 

400

 

±(0.8% 5

 

0.1

 

4k

 

±(0.8% 3

 

1

 

40k

 10

 

400k

 100

 

4M

 1k

 

40M

 

±(1.2% 5

 

10k

 

Overload protection: 250V RMS. 

NOTE: DO NOT input any voltage value at this range! 

3.2.7 Capacitance test 

Range Accuracy 

Resolution 

4nF 

±(5.0% 20

1pF 

40nF 

±(3.5%10

10pF 

400nF 100pF 

4uF 1nF 

40uF 10nF 

200uF 

±(5.0%10

100nF 

Overload protection: 250V RMS. 

NOTE: DO NOT input any voltage value at this range! 

3.2.8 Frequency test 

Range Accuracy 

Resolution 

100Hz 

±(0.5% 5

0.1Hz 

1kHz 0.1Hz 

10kHz 1Hz 

100kHz 10Hz 

1MHz 100Hz 

10MHz 1kHz 

Input sensitivity: >0.7V. 

Overload protection: 250V RMS. 

3.2.9 Temperature test 

 

 

Range Accuracy 

Resolution 

-40

750

)℃

 

400

±

1.0%+5

 

400

±

1.5%+15

1

 

Sensor: K type banana plug (Ni-Cr-Nickel silicon). 

NOTE: DO NOT input any voltage value at this range! 

3.2.10 Continuity Test 

Range Description 

Test 

Conditions 

 

Diode forward 

voltage drop 

Forward DC 

current is approx 

0.5mA, reverse 

voltage is 

approx 1.5V. 

 

When the resistance 

under test is less 

than 50±10

buzzer sounds 

continuously. 

Open circuit 

voltage: 0.5V 

Overload protection: 250V RMS. 

Note: DO NOT input any voltage value at this range! 

4. OPERATION 

4.1 PANEL DESCRIPTION                        (refer to picture 1) 

 Jaw: 0 to 600 AC current and NCV detecting 

device. 

 Clamp gunlock: Pressing the gunlock can turn on 

or off the clamp. 

 Hand protection: A safe design to protect users 

from touching the dangerous area. 

 Clamp light: Opening the clamp light can light up 

the tested area in the dark to prevent the danger. 

 NCV indicator: Detect the existence of the high voltage around to 

Summary of Contents for CM-2007

Page 1: ... To avoid electric shock and damage to the meter please disconnect the input signal and remove away the test leads before opening the cover Avoid the water leakage into the cover 2 2 3 DO NOT do any test if the cover hasn t been completely installed or screws not tightened up 2 2 4 DO NOT input the value higher than the limit at every range 2 2 5 DO NOT input any voltage at Resistance range 2 2 6 ...

Page 2: ... 100pF 4uF 1nF 40uF 10nF 200uF 5 0 reading 10 100nF Overload protection 250V RMS NOTE DO NOT input any voltage value at this range 3 2 8 Frequency test Range Accuracy Resolution 100Hz 0 5 reading 5 0 1Hz 1kHz 0 1Hz 10kHz 1Hz 100kHz 10Hz 1MHz 100Hz 10MHz 1kHz Input sensitivity 0 7V Overload protection 250V RMS 3 2 9 Temperature test Range Accuracy Resolution 40 750 400 1 0 5 400 1 5 15 1 Sensor K t...

Page 3: ...easurement inputs positive terminal red test lead COM input terminal Measurement inputs negative terminal black test lead Carrying belt 4 2 ACA MEASUREMENT 1 Turn the rotary switch to range 2 Open the clamp jaw and grip one wire and get the reading directly To get the most precise reading should user put the wire in the middle of the close jaw refer to the picture 3 Note 1 Firstly users should sel...

Page 4: ... the red test leads is for positive pole LCD displays capacitance value Note 1 Don t input voltage or current at Capacitance range 2 In order to assure the accuracy please press REL to clear the reading before testing 3 There is only the auto range mode under the capacitance range 4 The capacitor must be completely discharged before testing 5 The reading of 200uF range will take more than 15 secon...

Page 5: ...OLD key signal appears Replace battery Big error Value Replace battery The specifications are subject to changes without prior notice The content of this manual is regarded as correct If users find out any mistakes or omissions please kindly contact the manufacturer The manufacturer will not be responsible for accidents and damage caused by improper operations The functions described in this User ...

Reviews: