4.3.7. CAL key 

The 

CAL

 key permits to perform compensation on the temperature measurement due to a 

possible error of the input probe. Follow the below steps: 
 
1.  By pushing long (>3s) 

CAL

 key to enter in the calibration section. The below screen is 

shown at display: 

 

 

Fig. 3: Compensation of temperature probe 

 
2.  On the main display the real time value of the temperature is displayed while the 

measurement offset value is displayed on the secondary display. The default value is 
“0.0”. 

Do not modify this default value if any compensation is needed

 

3. Push the 

 or 

arrow keys respectively to increase or decrease the offset value. The 

maximum set values are: 

5°C

 or 

9°F

 with a 

0.1°C 

/

 0.1°F

 resolution 

4.  By pushing long (>3s) again 

CAL

 key to save the settings and quit this mode. The 

instrument keeps the set value after each shutdown  

 

4.3.8.  Disabling Auto Power OFF  

The instrument is provided with Auto Power OFF (APO) feature which automatically 
permits to switch it off after 15 minutes of idleness in order to preserve the internal battery. 
To disable this feature follows the below steps: 
 
1.  With instrument on, push and keep the 

HOLD/ENTER

 key 

2.  By pushing long (>1s) the 

ON/OFF

 key. The “APO OFF” message is shown at display 

for a while (see Fig. 4 – left part) and the “ “ symbol disappears from display 

3.  Repeat the same operation of the point 1 and 2 for enable again the function with “APO 

ON” message shown at display (see Fig. 4 – right part) or the feature is automatically 
restored after each shutdown 

 

 

 

Fig. 4: Disable/enable Auto Power OFF 

 
 

Summary of Contents for HTA103

Page 1: ... Copyright HT ITALIA 2017 Release EN 1 00 14 09 2017 ENGLISH User manual ...

Page 2: ... 3 2 UNIT key 5 4 3 3 JKT key 5 4 3 4 MAX and MIN key 5 4 3 5 HOLD ENTER key 5 4 3 6 and keys 5 4 3 7 CAL key 6 4 3 8 Disabling Auto Power OFF 6 5 OPERATING INSTRUCTIONS 7 5 1 Temperature measurement 7 6 MAINTENANCE 8 6 1 General 8 6 2 Battery replacement 8 6 3 Cleaning 8 6 4 End of life 8 7 TECHNICAL SPECIFICATIONS 9 7 1 Characteristics 9 7 1 1 General data 9 7 2 Environmental conditions 9 7 2 1 ...

Page 3: ... instrument only as specified in this manual otherwise the protection provided by the instrument may be impaired Only the accessories supplied with the instrument guarantee compliance with the safety standards They must be in good conditions and must be replaced if necessary with identical models Do not effect measurements beyond the limits specified in this manual Check that batteries have been p...

Page 4: ... in perfect conditions However it s advisable to make a rapid check in order to detect eventual damages which may have occurred in transit Should this be the case enter immediately the usual claims with the carrier Make sure that all the accessories listed in 7 3 are contained in the package In case of discrepancies contact the dealer In case of returning of the tester please keep to the instructi...

Page 5: ...ey 8 UNIT key 9 CAL key 10 JKT key 11 ON OFF key 12 Tripod hole 13 Battery cover Fig 1 Instrument description 4 2 DESCRIPTION OF DISPLAY CAPTION 1 Probe type indication 2 Auto Power OFF APO symbol 3 Low battery indication 4 Main display 5 Measurement unit 6 MAX function active 7 MIN function active 8 HOLD function active 9 Secondary display Fig 2 Display description ...

Page 6: ...e of external probe K J or T to be connected to the instrument Cyclically the K J or T settings are displayed 4 3 4 MAX and MIN key By pushing MAX or MIN key permits to detect the Maximum or Minimum values of measured temperature The secondary display see Fig 2 part 9 shows the symbol linked to the selected function The symbol MAX for maximum value and MIN for minimum value The MAX or MIN values a...

Page 7: ...or 9 F with a 0 1 C 0 1 F resolution 4 By pushing long 3s again CAL key to save the settings and quit this mode The instrument keeps the set value after each shutdown 4 3 8 Disabling Auto Power OFF The instrument is provided with Auto Power OFF APO feature which automatically permits to switch it off after 15 minutes of idleness in order to preserve the internal battery To disable this feature fol...

Page 8: ...wn at two displays 2 Select the type of probe see 4 3 3 and the measurement unit see 4 3 2 3 Insert the connectors of the probe on input terminal respecting the polarity and see Fig 1 part 1 shown on the instrument and on the probe connector 4 The real time value of the temperature is shown on the main display see Fig 5 Fig 5 Use of the instrument for temperature measurement 5 Press HOLD ENTER key...

Page 9: ...d trained technicians must perform this operation Remove all the probes from the devices under test before replacing the battery 1 Switch off the instrument 2 Remove the temperature probe 3 Remove the battery cover see Fig 1 part 13 4 Remove the battery from the battery fastener 5 Set the new battery into battery fastener and return it to the battery case 6 Replace the battery cover 7 Use the appr...

Page 10: ...1 F 1 rdg 1 8 F T 148 F 1 rdg 0 9 F T 148 F 7 1 1 General data Mechanical characteristics Dimensions Lx W x H 190 x 65 x 45mm 7 x 3 x 2in Weight including battery 235g 8 ounces Mechanical protection IP40 Supply Battery type 1x9V alkaline NEDA 1604 IEC 6F22 JIS 006P Low battery indication symbol is displayed Battery life ca 50h backlight ON ca 210h backlight OFF Auto Power Off after 15 minutes of i...

Page 11: ...de TK101 Battery Carrying bag User manual 7 3 2 Optional accessories Type K probe for air and gas Cod TK107 Type K probe for semi solid substances Cod TK108 Type K probe for liquid Cod TK109 Type K probe for surfaces Cod TK110 Type K probe for surfaces with a 90 lead Cod TK111 ...

Page 12: ...ue to improper packaging Faults due to servicing carried out by a person not approved by the company Faults due to modifications made without explicit authorization of our technical department Faults due to adaptation to a particular application not provided for by the definition of the equipment or by the instruction manual The contents of this manual cannot be reproduced in any form without our ...

Reviews: