Summary of Contents for HT9019

Page 1: ... Copyright HT ITALIA 2013 Release EN 2 00 18 04 2013 ENGLISH User manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 2: ...cription 7 4 2 1 H key 7 4 2 2 key 7 4 2 3 MODE key 7 4 2 4 MAX MIN key 7 4 2 5 REL key 7 4 2 6 PEAK key 7 4 3 Functions of rotary switch description 8 4 3 1 AC Voltage measurement 8 4 3 2 DC Voltage measurement 9 4 3 3 Resistance measurement and Continuity test 10 4 3 4 AC Current measurement 11 5 MAINTENANCE 12 5 1 General informations 12 5 2 Battery replacement 12 5 3 Cleaning 12 5 4 End of lif...

Page 3: ...ct use may damage the tester or its components High Voltage ranger electrical shock risk Double insulated instrument AC Voltage or Current DC Voltage or Current Ground reference 1 1 PRELIMINARY This apparatus has been designed for use in an environment of pollution degree 2 Indoor use It measures CURRENT and VOLTAGE on CAT IV 600V and CAT III 1000V plants For measurement categories please see 1 4 ...

Page 4: ... use Part 1 General requirements defines what measuring category usually called overvoltage category is On 6 7 4 Measuring circuits it says OMISSIS circuits are divided into the following measurement categories Measurement category IV is for measurements performed at the source of the low voltage installation Examples are electricity meters and measurements on primary overcurrent protection device...

Page 5: ... instruments provide only the value of the fundamental wave while True RMS instruments provide the value of the entire wave including harmonics within the passband of the instrument Accordingly if the same quantity is measured with both kinds of instruments the measured values are identical only if the wave is purely sinusoidal Should it be distorted True RMS instruments provide higher values than...

Page 6: ...The instrument is battery supplied One 1x9V battery type NEDA 1604 IEC 6F22 JIS 006P is included in the package Battery life about 200 hours When batteries are low the symbol appears on the display Replace them immediately following the instructions given in 5 2 The instrument disposes of the Auto Power OFF function not disable consisting in an automatic switching off 15 minutes of idleness 3 3 CA...

Page 7: ...EL PEAK function keys 7 LCD display 8 V input jack 9 COM input jack Fig 1 Instrument description 4 1 2 Alignment marks Put the conductor within the jaws on intersection of the indicated marks as much as possible see Fig 2 in order to meet the meter accuracy specifications CAPTION 1 Alignment marks 2 Conductor Fig 2 Alignment marks www GlobalTestSupply com Find Quality Products Online at sales Glob...

Page 8: ...e This mode is activated on each measurement except for continuity test This mode is disabled keeping pressed MAX MIN key or moving the rotary switch 4 2 5 REL key This key permits to perform the zeroing of value at display and a relative measurement of the parameter under test By pushing REL key the instrument takes the actual parameter s value as on offset and performs a relative measurement The...

Page 9: ...ng will be displayed 4 When O L symbol is displayed the result exceeds the maximum value that the instrument is capable of measuring 5 With meter switched on and rotary switch in each position move the integrated sensor close to an AC source The red LED turns on for detected voltage 100V 6 For HOLD MAX MIN and REL features please refer to 4 2 CAUTION Due to the high input impedance is possible tha...

Page 10: ... long ends of test leads to the desired circuit then reading will be displayed 4 When O L symbol is displayed the result is over the maximum value that the instrument is able to measure 5 For HOLD MAX MIN and REL features please refer to 4 2 CAUTION Due to the high input impedance is possible that the meter performs the zero of the display taking longer The possible blinking values shown at displa...

Page 11: ... test lead plug into COM jack Fig 5 3 Connect the two long ends of test leads to the desired circuit then reading will be displayed 4 Pushing MODE key and select continuity test The symbol is shown at display Buzzer emits sound if the measured resistance value is less about 60 5 When O L symbol is displayed the result under test exceeds the maximum value that the instrument is capable of measuring...

Page 12: ...ents Consider the notches on jaws as reference see Fig 2 The reading will be displayed 3 When O L symbol is displayed the result exceeds the maximum value that the instrument is capable of measuring Select an higher range 4 For HOLD MAX MIN REL and PEAK features please refer to 4 2 CAUTION The possible blinking values shown at display with open input terminals are not a problem of the meter and th...

Page 13: ...ed technicians must perform this operation Remove the test leads or the conductor under test before replacing the battery 1 Rotate the switch on OFF 2 Remove the test leads or the objects to be tested 3 Remove the screw from the battery cover and detach the battery cover from the bottom cover 4 Remove the battery 5 Replace the battery with a new one same type see 6 1 2 6 Replace the battery cover ...

Page 14: ...V 1000V 1V Integrated sensor for AC voltage detection LED turn on for phase earth voltage 100V 50 60Hz The meter emits a continuous sound with VDC 750V Resistance and Continuity test Autorange Range Resolution Uncertainty Buzzer Overload protection 600 0 0 1 1 0 rdg 5dgt 60 600VDC ACrms 6 000k 0 001k 60 00k 0 01k 600 0k 0 1k 6 000M 0 001M 60 00M 0 01M 2 0 rdg 10dgt Test current of continuity test ...

Page 15: ...ower OFF after 15 minutes of idleness not disabled Display Characteristics 4 LCD max 6000 counts decimal point unit symbol indication bargraph and backlight Sample rate 2 times sec Conversion mode TRMS 6 2 ENVIRONMENTAL CONDITIONS 6 2 1 Climatic conditions Reference temperature 18 C 28 C 65 F 83 F Operating temperature 5 40 C 41 F 104 F Operating humidity 75 RH Storage temperature 20 60 C 4 F 140 ...

Page 16: ...g carried out by a person not authorized by the company faults due to modifications made without explicit authorisation of our technical department faults due to adaptation to a particular application not provided for by the definition of the equipment or by the instruction manual The contents of this manual cannot be reproduced in any form without our authorization Our products are patented Our l...