F3000 

 

 

 

 

EN - 6

4.1.2. Opening/closing flexible clamp jaw 

 

 

CAUTION 

 

  Put the conductor as close as possible to the middle of the jaws  in order 

to meet the meter accuracy specifications (see Fig. 2) 

  When apply/remove the instrument on/from live conductors, if is not 

possible to de-energize the circuit on test, 

use appropriate security 

measures (e.g. protective gloves)

 to avoid dangerous electric shock for 

the operator and a possible damage of the instrument 

  Disconnect the main voltage or wear protective gloves before performing 

measurement  

  Put the hand below the LCD display during the measurement as indicated 

by the arrow in Fig. 2 

 

For opening/closing the flexible clamp jaw following the below steps: 
 
1.  Rotate of a quarter in the counterclockwise direction the fixing jaw connector to unlock 

the jaw (see Fig. 2) 

2.  Extract the flexible jaw and wrap around the conductor on test 
3.  Replace the jaw in the seat and rotate of a quarter in the clockwise direction the fixed 

jaw connector up to align the two arrows on the plastic case to lock the jaw (see Fig. 2) 
 

 

 

Fig. 2: Opening/closing flexible clamp jaw 

Summary of Contents for F3000

Page 1: ... Copyright HT ITALIA 2015 Release EN 1 02 23 06 2015 ENGLISH User manual ...

Page 2: ...e clamp jaw 6 4 2 Description of functions keys 7 4 2 1 HOLD key 7 4 2 2 Backlight key 7 4 2 3 Disabling the Auto Power OFF function 7 4 3 Operating instruction 8 4 3 1 AC Current measurement 8 5 MAINTENANCE 9 5 1 General information 9 5 2 Batteries replacement 9 5 3 Cleaning the instrument 9 5 4 End of life 9 6 TECHNICAL SPECIFICATIONS 10 6 1 Technical characteristics 10 6 1 1 Electrical specific...

Page 3: ...ng probes circuits etc Do not carry out any measurement in case you find anomalies in the instrument such as deformation breaks substance leaks absence of display on the screen etc In this manual and on the instrument the following symbols are used CAUTION observe the instructions given in this manual an improper use could damage the instrument or its components Double insulated meter AC current D...

Page 4: ...y is In 6 7 4 Measured circuits circuits are divided into the following measurement categories OMISSIS Measurement category IV is for measurements performed at the source of the low voltage installation Examples are electricity meters and measurements on primary overcurrent protection devices and ripple control units Measurement category III is for measurements performed on installations inside bu...

Page 5: ...l components indicated in 6 3 1 In case of discrepancy please contact the Dealer In case the instrument should be replaced please carefully follow the instructions given in 7 3 2 INSTRUMENT POWER SUPPLY The instrument is supplied with 2x1 5V AAA LR03 batteries included in the package In order to prevent compromising its charge the battery is not inserted in the instrument For batteries installatio...

Page 6: ...G INSTRUCTIONS 4 1 INSTRUMENT DESCRIPTION 4 1 1 Description of the controls CAPTION 1 Flexible clamp jaw 2 Opening closing jaw connector 3 LCD display 4 HOLD key 5 Backlight key 6 Battery cover Fig 1 Instrument description ...

Page 7: ...ent Disconnect the main voltage or wear protective gloves before performing measurement Put the hand below the LCD display during the measurement as indicated by the arrow in Fig 2 For opening closing the flexible clamp jaw following the below steps 1 Rotate of a quarter in the counterclockwise direction the fixing jaw connector to unlock the jaw see Fig 2 2 Extract the flexible jaw and wrap aroun...

Page 8: ...ight activation deactivation This light could help the operator reading the display while he s measuring in dark sites 4 2 3 Disabling the Auto Power OFF function In order to preserve internal batteries the instrument switches automatically off approximately 20 minutes after it was last used No symbol appears on the display when this function is active When the instrument must be used for long per...

Page 9: ...r and a possible damage of the instrument Disconnect the main voltage or wear protective gloves before performing measurement Put the hand below the LCD display during the measurement as indicated by the arrow in Fig 2 Correct Incorrect Fig 3 Use of the clamp for AC current measurement 1 Switch on the clamp by pressing the HOLD key 2 Open the flexible clamp jaw and insert the cable in the middle o...

Page 10: ...ns should perform this operation Before carrying out this operation make sure you have removed the cable being tested from inside the clamp jaw Fig 4 Batteries replacement 1 Switch off the clamp and remove the cable being tested from the clamp jaw 2 Rotate the fixing screw of the battery cover from the to the position remove the cover and extract the batteries see Fig 4 3 Fit two new batteries see...

Page 11: ...ollowed Error 2 0 rdg distance 15mm from center of jaws Error 2 5 rdg distance 25mm from center of jaws Error 3 0 rdg distance 35mm from center of jaws 6 1 1 Electrical specifications Conversion type TRMS Sample rate 4 times second Temperature coefficient 0 2x accuracy C 18 C o 28 C 6 1 2 Reference guideline Safety IEC EN61010 1 IEC EN61010 2 032 EMC IEC EN61326 1 Insulation double insulation Poll...

Page 12: ... C 40 C 86 F 104 F RH 75 40 C 50 C 104 F 122 F RH 45 Allowable relative humidity 80 RH Storage temperature 20 60 C 4 F 140 F Storage humidity 80 RH This instrument satisfies the requirements of Low Voltage Directive 2006 95 EEC LVD and of EMC Directive 2004 108 EEC 6 3 ACCESSORIES 6 3 1 Standard accessories User manual Carrying bag Batteries not fitted ...

Page 13: ...anufacturer s explicit authorization Use not provided for in the instrument s specifications or in the instruction manual The content of this manual cannot be reproduced in any form without the manufacturer s authorization Our products are patented and our trademarks are registered The manufacturer reserves the right to make changes in the specifications and prices if this is due to improvements i...