5.2. REPLACEMENT OF INTERNAL FUSES
Fig. 32: Replacement of internal fuses
1. Position the rotary switch to OFF and remove the cables from the input terminals.
2. Turn the fastening screw of the battery compartment cover from position
“
”
to position
“
”
and remove it (see Fig. 32)
3. Remove the damaged fuse and insert a new fuse of the same type (see § 6.1.2).
4. Restore the battery compartment cover into place and turn the fastening screw from
position
“
”
to position
“
”
.
5.3. CLEANING THE INSTRUMENT
Use a soft and dry cloth to clean the instrument. Never use wet cloths, solvents, water, etc.
5.4. END OF LIFE