TROUBLESHOOTING GUIDE
PROBLEM
PROBLEM CAUSE
SOLUTION
CHAIR WILL NOT OPERATE.
1) POWER CORD NOT PLUGGED IN.
2) NO AC POWER AT WALL RECEPTACLE.
3) BLOWN LINE FUSE.
1) CHECK ALL WIRING CONNECTIONS.
2) RESTORE OUTLET POWER.
3) REPLACE FUSE.
OUTLET POWER OK, BUT NO
BEEP IS HEARD WHEN POWER
IS APPLIED.
1) BLOWN LINE FUSE.
2) LOOSE WIRING IN BASE.
3) OTHER.
1) REPLACE FUSE.
2) CHECK ALL WIRING CONNECTIONS.
3) CONTACT FACTORY
FOOTSWITCH FUNCTIONS NOT
OPERATING.
1) FOOTSWITCH CORD NOT CONNECTED.
1) CHECK ALL INTERNAL AND
EXTERNAL FOOTSWITCH WIRING
BACKSWITCH FUNCTIONS NOT
OPERATING.
1) DAMAGED OR DISCONNECTED WIRES
IN CHAIR BACK, UNDER CHAIR SEAT,
OR IN CHAIR BASE
1) CHECK ALL WIRING CONNECTIONS.
CHAIR BASE WILL NOT STAY UP 1) SOLENOID VALVE NOT CLOSED.
(DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER
1) REPAIR OR REPLACE SOLENOID
VALVE.
2) CHECK FOR SWITCHES STUCK
"CLOSED" .
3) CONTACT FACTORY.
CHAIR BASE WILL NOT LOWER. 1) NO POWER TO PUMP MOTOR ASS'Y.
2) SOLENOID COIL BURNED-OUT
3) OTHER
1) CHECK FOR VOLTAGE.
2) REPLACE SOLENOID COIL.
3) CONTACT FACTORY.
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR DOES NOT RUN.
1) NO POWER TO PUMP MOTOR ASS'Y
2) UP-LIMIT SWITCH STUCK "OPEN".
3) PUMP MOTOR BURNED-OUT.
4) OTHER
1) CHECK FOR VOLTAGE.
2) RE-ADJUST OR REPLACE SWITCH.
3) REPLACE PUMP MOTOR ASS'Y.
4) CONTACT FACTORY.
CHAIR BASE WILL NOT RAISE.
PUMP MOTOR RUNS.
1) SOLENOID VALVE NOT CLOSED.
(DIRT IN VALVE.)
2) CONSTANT POWER APPLIED TO
SOLENOID.
3) OTHER.
1) REPAIR OR REPLACE SOLENOID
VALVE.
2) CHECK FOR SWITCHES STUCK
"CLOSED".
3) CONTACT FACTORY.
BASE "STUTTERS" DURING
ASCENT.
1) HYDRAULIC FLUID LEVEL IS LOW.
1) CONTACT DEALER OR FACTORY
FOR INFORMATION ON ADDING OIL.
CHAIR BASE WILL NOT RAISE
OR LOWER FROM FLOOR UNIT.
1) CHAIR CONTROL CABLE NOT PLUGGED
IN.
2) BLOWN FUSE IN CHAIR BASE.
1) CHECK CABLE CONNECTIONS AT
STAND CHAIR
2) REPLACE FUSE.
CHAIR TOP "WOBBLES" ON
CHAIR BASE
1) LOCK NUT IS LOOSE.
(UNDER SEAT BOARD.)
1) TIGHTEN LOCK NUT.
CHAIR BACK RECLINES FULLY
WITHOUT HANDLE BEING
ACTIVATED.
1) BACK SPRING FAILURE
1) ADJUST GAS SPRING
2) REPLACE GAS SPRING
CHAIR BACK SPONGY WHEN
PATIENT PUTS PRESSURE ON
CHAIR BACK.
1) BACK SPRING FAILURE
1) REPLACE GAS SPRING
IN-6200
12
Summary of Contents for 6200
Page 1: ...OWNERS MANUAL INSTALLATION AND OPERATING INSTRUCTIONS MODEL 6200 LOW VOLTAGE CONTROL CHAIR ...
Page 19: ...IN 6200 19 NOTES ...
Page 24: ...FIGURE 11 18 22 HEADREST ASSEMBLY IN 6200 24 1 3 5 6 9 9 2 11 8 7 13 12 14 15 16 17 4 10 18 ...
Page 26: ...FIGURE 12 10 HEADREST ASSEMBLY IN 6200 26 3 5 4 2 7 6 11 1 8 12 10 9 4 4 4 ...
Page 28: ...1 14 4 13 16 17 12 2 11 3 8 10 17 9 5 15 6 7 4 4 18 FIGURE 13 11 HEADREST ASSEMBLY IN 6200 28 ...
Page 34: ... NOTES IN 6200 34 ...