Calibrating the spectrophotometer
Performing a full calibration
Before using the spectrophotometer for the first time, and once a month thereafter,
perform a full calibration as follows:
1
From the
Utilities
menu, click
Calibrate Spectrophotometer
, and
Full Calibration
.
2
Follow the instructions in the series of message windows that appear, clicking
OK
after each message window.
Note
For the Black Calibration test, make sure that you measure the dark area until the four vibrating
measurements end completely. This takes approximately 15 seconds.
If the
Black calibration successful
and the
White calibration
successful
messages appear on the screen, the full calibration was successful. If
not, repeat this procedure one more time before calling the customer engineer.
Performing the daily calibration
Calibrate the spectrophotometer once a day.
1
From the
Utilities
menu, click
Calibrate Spectrophotometer
, and
Frequent
Calibration
.
2
Follow the instructions in the window that appears.
Note
For the White Calibration test, make sure that you measure the white reference area until the
single vibrating measurement ends completely. This takes approximately two seconds.
If the
Calibration successful
message appears on the screen, the daily
calibration was successful. If not, repeat this procedure one more time before calling
the customer engineer.
Summary of Contents for IndiChrome
Page 1: ...HP IndiChrome Ink Mixing System User guide ...
Page 2: ...HP IndiChrome Ink Mixing System User guide ...
Page 7: ...6 Contents ENWW ...
Page 63: ...62 Chapter 4 General procedures ENWW ...
Page 92: ...ENWW Program updates 91 ...
Page 93: ...92 Chapter 7 Help ENWW ...
Page 103: ...EN 102 ...