prog.book : ch9sens.fm 52 Mon Feb 15 09:24:37 1999
SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
9-52
HP E4418B/E4419B Programming Guide
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO <Boolean>
This command is used with the HP E-series power sensors to enable and
disable autoranging. When autoranging is
ON
, the power meter selects the
best measuring range for the measurement. When autoranging is set to
OFF
, the power meter remains in the currently set range.
The
[SENSe[1]]|SENSe2:POWer:AC:RANGe
command disables
autoranging.
If
INITiate:CONTinuous
is set to
ON
and
TRIGger:SOURce
is set to
IMMediate
, the range will track the input power if
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO
is
ON
.
If the power meter is not making measurements then autoranging only
occurs when the power meter is triggered.
Syntax
Example
Reset Condition
On reset, autoranging is enabled.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO?
The query enters a
1
or
0
into the output buffer indicating the status of
autoranging.
•
1
is returned when autoranging is enabled.
•
0
is returned when autoranging is disabled.
POW:AC:RANG:AUTO 0
This command disables
autoranging.
SENS
1
:AC
POW
SENS2
:
:AUTO
0|OFF
1|ON
Space
?
:RANG
Summary of Contents for E4418B
Page 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Page 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Page 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Page 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Page 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Page 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Page 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Page 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Page 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Page 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Page 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Page 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Page 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Page 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Page 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Page 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...