4.
Open the channel output relay by sending the
OUTPut
n OFF
command, where
n
is the channel number.
5.
Change the connections to the next channel and repeat steps 2, 3,
and 4.
6.
Repeat steps 2, 3, 4, and 5 for each channel.
DC Voltage
Compliance Current
This test checks the compliance current in dc voltage.
1.
For each channel in the module, connect the 600
Ω
resistor and the
DMM as shown in Figure 4-2. Channel 1 connections are shown in
the figure. Connect the other channels in a similar manner. Set the
DMM to measure dc voltage.
2.
Set the channel to 12.00 V. Use the
APPLy
n:VOLTage 12.00
command, where
n
is the channel
number.
3.
Measure the output voltage with the DMM, enter it in the table
beginning on page 46, and compare the measured voltage to the limits
shown in the table. The compliance current is inferred from the
voltage across the resistor.
4.
Open the channel output relay by sending the
OUTPut
n OFF
command, where
n
is the channel number.
5.
Change the connections to the next channel and repeat steps 2, 3,
and 4.
6.
Repeat steps 2, 3, 4, and 5 for each channel.
Figure 4-2. DC Voltage Output Connections (Channel 1)
Chapter 4
Verification Tests 35]
Summary of Contents for E1418A
Page 4: ...Notes 4 HP E1418A 8 16 Channel D A Converter Service Manual Contents ...
Page 8: ...Notes 8 HP E1418A D A Converter Module Service Manual ...
Page 58: ...Notes 58 Verification Tests Chapter 4 ...
Page 66: ...Figure 5 5 Internal Resistor Adjustment Connections 66 Adjustments Chapter 5 ...
Page 81: ...Figure 6 1 8 Channel Disassembly Chapter 6 Service 81 ...
Page 82: ...Figure 6 2 16 Channel Disassembly 82 Service Chapter 6 ...
Page 84: ...Notes 84 Service Chapter 6 ...
Page 90: ...Figure 7 1 HP E1418A Replaceable Parts 90 Replaceable Parts Chapter 7 ...
Page 91: ...Figure 7 2 Common Terminal Module Replaceable Parts Chapter 7 Replaceable Parts 91 ...