Maintaining the Defibrillator
Cleaning the Recorder Printhead
8-2
3
Place a new roll of thermal paper (HP 40457C/D) in the recorder so that the
paper unrolls from the top of the roll and the grid faces down as it comes out of
the recorder.
4
Pull the end of the paper past the recorder platen.
5
While holding the recorder door open, (to the right) press the platen down over
the paper.
6
Allow the door to close over the platen roller.
8
Cleaning the Recorder Printhead
If the ECG strip has light or varying density printing, clean the printhead to remove
any possible buildup of paper residue.
To clean the printhead, perform the following steps.
1
Slide the recorder door to the right of the defibrillator. The paper platen will tilt
up.
2
Remove the paper roll.
3
Clean the printhead surface (above the brush) with rubbing alcohol and a cotton
swab.
4
Re-install the paper roll.
5
While holding the recorder door open, (to the right) press the platen down over
the paper.
6
Allow the door to close over the platen roller.