HP 85325A Operating and Service Manual
Service
Recommended Test Equipment
Band-Dependent
Equipment
Table 6-2 shows equipment that is mm-wave band dependent. Fixed
terminations, variable attenuators, and directional couplers are only usable
in one band. Therefore, if your system has more than one mm-wave band,
you must have a termination, variable attenuator, and directional coupler for
each band.
Table 6-2
Recommended Test Equipment (Banded)
Item
Critical Specification
Recommended Model or Part
Number
Use
1
Power Meter
No substitute
Anritsu ML83A (for V or U bands)
Power Sensors
2
Return loss > 14 dB
R band: HP R8486A
Q band: HP Q8486A
U band: Anritsu ML715A
V band: Anritsu ML716A
W band: HP W8486A
P, T
Fixed Termination
2
Return loss > 30 dB
HP x910A
3
P, T
Variable Vane Attenuator
2
0 to 40 dB attenuation
HP x382A
3
V band, Millitech DRA-15
W band, Millitech DRA-10
P
Directional Coupler
2
Coupling factor 10 dB
HP x752C
3
P, T
1. I=installation; O = operation; P = performance verification; T = troubleshooting and repair
2. You must have one of these devices for each mm-wave band
3. Where x stands for the band R, Q, U, V, or W.