Servicing
the
HP
83447A
P
erformance
T
ests
12.
Change
the
clock
frequency
to
155
MHz.
Measure
the
jitter
again,
as
above:
at
wavelength
1310
nm:
T
otal
RMS
jitter
at
155
Mbps
(129
ps,
max)
at
wavelength
1550
nm:
T
otal
RMS
jitter
at
155
Mbps
(129
ps,
max)
13.
Change
the
clock
frequency
to
50.0
MHz.
Measure
the
jitter
again,
as
above:
at
wavelength
1310
nm:
T
otal
RMS
jitter
at
50
Mbps
(400
ps,
max)
at
wavelength
1550
nm:
T
otal
RMS
jitter
at
50
Mbps
(400
ps,
max)