F
Device-Specic
Errors
These
are
error
messages
in
the
range
-300
to
-399,
or
between
1
and
32767.
They
show
that
an
error
has
been
detected
that
is
specic
to
the
operation
of
the
polarization
controller
.
A
device-specic
error
is
signaled
by
the
device-specic
error
bit
(bit
3)
in
the
event
status
register
.
-300
Device-specic
error
.
This
shows
that
a
device-specic
error
has
occurred.
No
more
specic
information
is
available
.
-310
System
error
.
An
instrument
system
error
has
occurred.
-311
Memory
error
.
A
memory
error
has
been
detected.
-314
Save/recall
memory
lost.
The
nonvolatile
data
saved
by
the
*SAV
command
has
been
lost.
-315
Conguration
memory
lost.
The
nonvolatile
conguration
data
saved
by
the
instrument
has
been
lost.
-330
Self-test
failed.
Further
information
about
the
self-test
failure
is
available
by
using
*TST? .
-350
Queue
overow
.
The
error
queue
has
overown.
This
error
is
written
to
the
last
position
in
the
queue
,
no
further
errors
are
recorded.
F-6
Error