D
F
or
option
#020
rst
splice
a
HP
81102BC
patchcord
to
the
pigtail
of
the
P
ol-port.
Make
sure
that
all
instruments
have
warmed
up
.
Fix
all
cables
with
tape
so
that
they
won't
move
during
measurements
.
3.
Set
up
the
8153A.
a.
Zero
the
8153A:
press
4
ZERO
5 .
b.
Set
the
8153A
to
dBm:
press
4
dBm/W
5
until
display
shows
dBm.
c.
Set
the
8153A
to
wavelength
=
1470
nm:
press
4
pa
ram
5
until
is
shown,
set
to
1470
using
the
Modify
keys
.
d.
Set
the
8153A
sample
time
to
50
ms:
press
4
pa
ram
5
until
you
get
T,
use
the
Modify
keys
,
until
display
reads
50ms
.
e.
Set
8153A
to
datalogging:
i.
Press
4
Mo
de
5
to
get
MENU.
ii.
Press
4
Reco rd
5
until
you
get
LOGGING.
iii.
Press
4
Edit
5
to
get
SAMPLES.
iv
.
Press
4
Mo
dify
5
cursor
until
display
reads
500.
v
.
Press
4
Next
5
until
you
get
START.
vi.
Press
4
Mo
dify
5
cursor
until
you
get
IMMEDIAT.
vii.
Press
4
Edit
5
to
get
LOGGING.
4.
Set
up
the
8168C.
a.
Press
4
W
avelength
5 ,
press
4
Edit
5 ,
type
1470
on
the
numeric
keypad,
press
4
Enter
5 .
b.
Press
4
Output
Pow
er
5 ,
press
4
Edit
5
type
100
on
the
numeric
keypad,
press
4
Enter
5 .
c.
A
ctivate
the
Optical
Output.
5.
Set
up
the
8169A
(DUT)
a.
Set
the
polarizing
lter
for
maximum
transmission
(see
\Setting
the
P
osition
of
the
P
olarizing
Filter"
in
Chapter
2).
b.
Press
4
Sphere
5 ,
select
Slow
with
the
Modify
knob
and
press
4
Enter
5 .
P
erformance
T
est
D-3