SVC 10- 6
suitable ground. Failure to heed this caution
may result in damage to the instrument.
9. Slide the rear panel away from the back of the instrument, taking care not to damage the rear
panel connector PCB ribbon cable.
10. If the rear panel connector PCB is to be replaced, remove it from the rear panel.
11. Free and remove the ribbon cable from the rear panel connector PCB by releasing the locking
tabs (one on either side of the connector receptacle) and pulling the cable straight out.
12. If the rear panel connector PCB ribbon cable is
to be replaced, free and remove the cable from
the connector on the communications PCB by
releasing the locking tabs (one on either side of
the connector receptacle) and pulling the cable
straight out.
13. If the ribbon cable is to be replaced, slide the
old ribbon cable out of the opening in the elec-
tronics carrier.
14. If the ribbon cable is being replaced, slide the
new ribbon cable into the opening in the elec-
tronics carrier.
15. If the ribbon cable is being replaced, install the
new ribbon cable into the connector receptacle
on the communications PCB. (When the con-
nector is properly seated, the locking tabs on
the connector receptacle will wrap around the
edge of the connector, locking it in place.)
16. Install the ribbon cable into the connector receptacle on the rear panel connector PCB. (When
the connector is properly seated, the locking tabs on the connector receptacle will wrap around
the edge of the connector, locking it in place.)
17. If a new rear panel connector PCB is being installed, secure it to the rear panel using three
screws.