Chapter 2 Service
About the HP 53131A/132A Calibration Menu
2-12
Assembly-Level Service Guide
2
Equipment and Counter Setup: See Figure 2-1.
Figure 2-1. Calibration Setup for CAL: TI FINE?
Out of HP 8130A, into HP 59992A J06 Calibrator Signal:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V
Low:
−
0.50 V
Input Mode: Normal
Timebase: Any external timebase you provide is ignored during
calibration.
Procedure: From the front-panel calibration menu, one keypress
invokes the calibration. You are prompted to press buttons on the
HP 59992A J06 Calibrator for each part of a four-part calibration.
Note that CAL: TI FINE? requires the completion of four steps in
order.
HP 53131A/132A
Counter
1
2
3
4
In
A
B
HP 59992A J06
Time Interval Calibrator
Start
Stop
HP 8130A
Pulse Generator
Output
Summary of Contents for 53131A
Page 1: ...Assembly Level Service Guide HP 53131A 132A 181A 225 MHz Counter ...
Page 2: ......
Page 14: ...Contents xii Assembly Level Service Guide ...
Page 22: ...Preface xx Assembly Level Service Guide ...
Page 23: ...1 Performance Tests Verifying Specifications ...
Page 97: ...2 Service ...
Page 149: ...3 Replacing Assemblies Disassembly and Reassembly ...
Page 173: ...4 Retrofitting Options ...
Page 191: ...5 Replaceable Parts ...
Page 201: ...6 Backdating Manual Changes ...
Page 218: ...Chapter 6 Backdating Backdating Specifications 6 18 Assembly Level Service Guide 6 ...
Page 219: ...7 HP 53131A 132A Specifications ...
Page 235: ...8 HP 53181A Specifications ...
Page 246: ...Chapter 8 HP 53181A Specifications General Information 8 12 Assembly Level Service Guide 8 ...
Page 252: ......
Page 254: ... Manual Part Number 53131 90039 Printed in U S A MARCH 1999 ...