background image

Model  456A 

Section  IV 

Pa ragraphs  4-1  [0 4-5 

SECTI ON 

IV 

MAINTENANCE 

4-1 . 

G ENERAL. 

The Mode1 456A should require very littl e maintenance. 
It  is  built  w ith  etched  circuit wi ring  and  transistors 
which  should  ens ure long;  trouble-free  life .  However, 

should  trouble  occur  special  care  must  be  taken  in 
s ervicing  to  a void  damage  to  the  (ransistors  or  the 
etched  circuit board. 

REPLAC ING  TRA NSISTORS. 

The  transistors  are 

soldered  in:  substitution  of  transistors  should  not be 

resorted  to  unless  there  is some  indication  that they 

are  fa ulty.  Unless  extreme care is taken when  solder­

ing,  the  transistor  may  be damaged.  See figu re 

4-1. 

CAUTION 

Be  careful  not  to  short  voltages across  th e 

transistors . 

Small  bias  changes  may  ruin 

a  transistor  due  to  excessive  dissipation. 
Be  sure  to  turn  the  unit off  before  doing any 
soldering.  A  smaUleakage current  from the 

iron  applied  at  the  input  may  exceed  ratings 

on  the  transistors  at the  output. 

Components  with in 

� 

i nstruments  a re  conserva tively 

operated  to  provide  maximum  ins trument  reliability. 

In  spi te  of  this ,   parts  within  an  ins trument  may  fail. 

Usually,  the  instrument must  be immedia tely repaired 
with  a  minimum  of  "down  time". 

A  s ystematic 

approach  can  greatly  simplify  and  thereby  speed  up 
the  repair. 

Specifications  for  the  Model  456A 

AC 

Current  Probe 

are given  in  table  1-1 .   The procedures in  this  manual 

give  additi onal tests  and  the data  they  contain  are not 

to  be  considered  as  s pecifica tions . 

Your  Hewlett-Packard  representa tive  maintains com­
plete  facilities  and  specially  trained  personnel  to 
a ssist  you  with  any  engineering,  applictttion,  tests, 
or  repair  problems  you  may  have  with  Hewlett­

Packard  ins truments . 

4-2. 

CABIN E T  REMO V Al. 

CAUTION 

Remove  power  cord  from  receptacle if unit 
is  equipped  with  ac  s upply. 

The  instrument  may  be  taken  from  the  cabinet  by 
remOVing the two sc rews  at  the  rear of the  instrument 

and  pulling  the cabinet  loose  from  the front  panel  a nd 
side  frames. 

4-3. 

O P �R AT IO N   O N   230  VO LT  AC. 

The  Model 456A  is  normally wired for  1 15  volt oper­
ation  unless  otherwise  specified.  It  can  be  quickly 

and  easily  converted  to  operate  from  a  nominal line 

00395-1 

voltage  of  either  11 5  or  230  volts  and  a  frequency  of 
50  to  1 000  cps .  To  convert  for  230-volt opera tion 

remove  the  short  across  R 90.  Tu rn  the  instrument 

off before doing any  soldering. 

4- 4 .  

C HEC KI N G 

THE 

BATT E R IES. 

Whenever  trouble  is  encountered  check the  voltage a t  
the  BATTERY  TEST  terminals.  If  this  VOltage  is 

below  7  vol ts  replace  all  of  the  batteries .  The  B AT­

TER Y  TEST  terminals  are  across  only  pa rt  of  the 
batteries .  F or  this  reason all  of  the  batteries  should 
be  replaced  if  the  voltage  is  below  7  vol ts .   If  some 
of  the  ba tteries  are  depleted  the  remaining  batteries 
are  proba bly  low also.  In addition, the batteries  acrOSs 

the  BATTERY 

TEST 

jacks  could 

be 

fresh  while  �he 

other  battery  could  be  depleted.  If  trouble  is  en­
countered,  remove  the  cabinet  and  test  the  voltage 

of BT3  also.  Th is  voltage  should be at least 4.7  volts . 

Discard  battery  if  lower. 

WARNING 

Mercury  ba tteries  generate hydrogen  gas  at 

the  end of thei r  l ife  or when they a re  shorted . 
Hydrogen  gas  is  highly  explos ive and  may  be 

exploded  by  the  heat of  the  battery.  Observe 
the following precautions when us ing mercury 
ba tteries. 

1)  Never  discha rge  a  mercury-cell  battery 

after  its  voltage  falls below 7D% of its nominal 

voltage,  or  when  it fa ils  to operate the equip­
ment in which  i t  is  used . 

2)  Never  place  a  direct  short  ac ross  a 
mercury cell  battery. 

3) 

Never  leave  the 

POW 

ER switch 

ON 

w hen 

the  equipment  is  not  in  use,  or  a fter  the 
battery  fails  to  operate  the  equipment. 

4) 

Never retain  exhaus ted mercury -cell bat­

teries .  Discard  dead  batteries  as  soon  as 
possible.  Discard  in ga rbage,  NOT  in  was te 
basket.  if  batterv is  incinerated.  in  addition 
to  the  explosive  haza rd,  mercury  vapor  is 
EXTREMEL Y  TOXIC. 

DO 

NOT INCINERATE. 

5)  Store  spare  mercury-cell  batteries  in  a 

cool,  adequately  ventilated  area . 

4-5 .  

RE P LAC I N G  THE  BAT TER I E S .  

Whenever  the  voltage  at the  BATTERY  TEST  termi­
nals  falls  below  7  volts  or  if  the batteries  are  sus­
pected,  replace  the  batteries.  Turn  (he  ins trument 

OFF. 

Remove  the  cabinet as  instructed  in  paragraph 

4-2.  Remove  and  disca rd  all  ba tteries.  Replace  the 

batteries  with  one  Mallory  Type 

TR 

234  battery  or 

4-1 

Summary of Contents for 456A

Page 1: ...56A AC CURRENT PROBE SERIALS PREFIXED 103 PRINT O 3 61 I 00395 1 ...

Page 2: ...00395 1 OPERATING AND SERVICING MANUAL MODEL 456A SER IALS PREFIXED 103 AC CURRENT PROBE Copy ht HEWLETf ACK UD COM ANY 1961 ISOI AGE Mill ROAD AlO AlIO CALIFORNIA U S A Printed MAR 1961 ...

Page 3: ... 13 Replacing the Probe Cable REPLACEABLE PARTS 5 1 5 2 5 3 Introduction Table of Replaceable Parts Ordering Information 4 L 4 1 4 L 4 1 4 1 4 3 4 3 4 3 4 3 4 4 4 4 4 4 4 4 5 1 5 L 5 1 5 1 LIST O F IllUSTRATIONS Number Title L l General A rrangement of Model 456A 1 2 Model 456A AC Current Probe 1 3 Typical Change in F requency Response 2 1 2 2 2 3 2 4 2 5 with Load Capacitance Probe used with Mcxi...

Page 4: ... probe Less than 50 milliohms in series with 0 05 h This is approximately the induc tance of 1 1 2 inches of hookup wire Approximately 4 pf added from wire to ground For 1 2 amp input at least 50 db down For 10 rna input at least 70 db down Less than 50 a rms 100 a when ac powered 220 ohms at 1 kc Approximately 1 volt dc component Should work into load of not less than 100 000 ohms shunted by appr...

Page 5: ...or The probe operates with an accompanying small ampli fier to convert the ac current being measured to a pro portional voltage The current to voltage conversion factor is 1 millivolt output for 1 milliampere flowing th rough the probe Current readings can thus be taken directly from the voltage calibrations on the voltmeter or oscilloscope Table 1 2 is a list of instruments w hich can be used wit...

Page 6: ... The Model 456A is available either as a battery powered or ac powered instrument The AC Power Supply srock 456A 95B fits into the space nor mally occupied by the battery supply and takes the place of it electrically The ac power supply must be ordered separately When the 456A is powered with its own internal bat teries it is independent of power lines It may be used in the field with battery oper...

Page 7: ...ribed in text 00395 1 In this manner currents even below 10 microamperes can be measured 2 3 OPERAT IN G T HE PRO BE The probe jaws may be operated with one hand While holding the probe handle in the palm squeeze the flanges together with the fingers This action will open the jaws Move the probe to the vicinity of the insulated wire being measu red and slowly release the flanges until the jaws fit...

Page 8: ...by squeezing flanges on probe together 3 Connect ground clip to instrument under test if necessary see paragraph 2 5E Figure 2 2 2 2 Model 456A 4 t 1I t f RO 4 Read current being measured on meter or oscil loscope read millivolts as milliamperes S Note direction of arrow on probe Conventional current flow in direction of arrow gives positive going output voltage Opera tion 00395 1 ...

Page 9: ...n the signal being measured should preferably be kept below 1 2ampere Direct current below that value will have no effect on the measurement C ALTERNATING CURRENT FIELDS The probe is shielded against external ac magnetic and electric fields Extremely strong fields may cause an erroneous reading If a strong ac magnetic field 00395 1 is suspected rotate the empty probe head with the jaws closed If t...

Page 10: ...u may choose which side of the circuit to measure the current See figure 2 5 Note in figure 2 5 the probe measures the current flowing in the circuit on the same side as the wide side of the arrow on the probe Occasionally at high frequencies above 15 me a greatly magnified voltage effect exists at a particular frequency This effect is caused by a resonance be tween excessive external ground lead ...

Page 11: ... to the tube output capacity hp 456A PROBE HEAD Section II Paragraph 2 6 Cont d The probe is valuable in a number of other current equalizing applications EspeciaUy common among these are equalizing the input and output currents in push pull and balanced circuits in both transistor and tube applications Figure 2 6 Bucking OUt Capacitive Component of Plate Current 00395 1 2 5 2 6 ...

Page 12: ...results in a better frequency response lower input impedance dc bias stability lower distortion etc le 4 RD Figure 3 1 Equivalent Circuit 00395 1 The equivalent circuit can be thought of as shown in figure 3 1 where CL is the current gain of the tran sistor The transistors have internal resistances as shown Emitter resistance re is small but relevant to this discussion while the collector resistan...

Page 13: ...lowed extensive damage to the etched circuit board will result 1 Apply heat sparingly to lead of component to be replaced If lead of component passes through an eyelet in the circuit board apply heat on com ponent side of board If lead of component does not pass through an eyelet apply heat to con dUctor side of board 3 Bend clean tinned leads on new part and care fully insert through eyelets or h...

Page 14: ...tion unless otherwise specified It can be quickly and easily converted to operate from a nominal line 00395 1 voltage of either 115 or 230 volts and a frequency of 50 to 1000 cps To convert for 230 volt operation remove the short across R90 Turn the instrument off before doing any soldering 4 4 C HECKIN G THE BATTERIES Whenever trouble is encountered check the voltage at the BATTERY TEST terminals...

Page 15: ...il400D H L reads exactly 0 1 volt Model 456A 456A 5 Connect the OUTPUT cable of the 456A to the input of the 400D H L in place of the 470C Leave the wires still connected to the 470C 6 Read 400D H L Reading should be within sum of specifications of the 470C and 456A 7 Repeat above steps with 650A set at other fre quencies up to 100 kc If the 456A calibration or frequency response is poorer than te...

Page 16: ...illoscope should be less than 75 Jl volts 100 Jl volts with ac powered model If not check C94 power supply capacitor if ac powered or try replacing CRI Q1 or Q2 Check with 456A POWER switch off to see if ground loops exist If they do see paragraph 2 50 B CALIBRATION A ND FREQUENCY RESPONSE Perform the procedure illustrated in figure 4 2 Note that this test is used only for rough testing with readi...

Page 17: ...o the semi conductor Table 4 1 lists the tests which should be made after replacing the components 4 11 CALIBRATION Perform the procedure in figure 4 2 at 1 kc Adjust R9 Gain Set Potentiometer so that the voltage out is exactly 0 1 volt millivolts output equals milli amperes input If you can t reach calibration check R7 9 10 and 12 and the probe 4 4 Model 456A 4 12 FREQUEN CY R ESPO N S E Perform ...

Page 18: ...TEXT LOCKWASH ER l 2190 0012 PROBE CABLE ASSEMBLY COMPLETELY ASSE MBLED 4 5 6A 2 IA B PROBE G R OU N D LEAD ASSEMBLY CONTAI NS 2 2 OHM RESI STOR 456 A 2I A 10 PROBE F L CENTER SECTION COIL SP R I NG TO C3 i FRONT PANEL I I O F I NSTRUMENT LI INTERNAL LOCKWASH E R Figure 4 3 Exploded View of P robe Assembly Section IV Figure 4 3 00395 1 4 5 ...

Page 19: ...T 1101 OPEN leOV K T9 II o 11 0 Y 111 0 f nUIA t1191 r Y IV WK 011 1110 wlff v NOMINAL 11 0 2 J TO ull o RII 2 7 BL f WHT A 860 C90 4OI1F O OI f e5 O OI F e6 10UF R7 3300 AC POWER SUPPLY Rn 90 10K 40 IIf R93 6100 c 1 Q Q90 2N650 REGUL ATOR 131 BATTERY SUPPLY IIC IJLATC D r o 811 lIT 1 T lIT 2 f oOV BTl C O S 4V n 14 R IZ 100 U1 Figure 4 4 Model 456A AC Current Probe R9 ICO e93 lOll RIO 100 RI lao ...

Page 20: ......

Page 21: ... 3 O R DERI N G IN FORMAT I O N To order a replacement pa rt address order or in quiry either to your authorized sales office or to CUSTOMER SERVICE Hewlett Packard Company 395 Page Mill Road Palo Alto California or in Western Europe to Hewlett Packard S A Rue d u Vieux Billa rd No 1 Geneva Switzerland Specify the following information on a part 1 Model and serial number of the instrument Be s u r...

Page 22: ...Diode germanium IN38A Diode silicon Connector female red Connector female black Connector male output Not assigned Cable power Transistor 2N1 516 0C1 70 ot assigned rransistor 2N6 50 esistor fixed composition 24 ohms 5 1 W esistor fixed composition 22 ohms Part of Probe Ground Lead Assembly Component not separately replaceable Mfr 1 Stock No 9 609 5 01 50 0008 5 6289 0 180 0050 5 62 89 0180 0061 5...

Page 23: ...igned Resistor fixed composition 1 5 000 ohms 10 2 W Resistor fixed composition 680 oluns 10 1 2 W Resistor fixed composition 10 000 ohms 1 0 1 2 W Resistor fixed composition 6800 oluns 10 1 2 W Resistor fixed deposited carbon 4000 ohms 1 1 2 W Resistor fixed depOSited carbon 2 2 50 0luns 1 1 2 W Switch slide DPDT Not assigned Transformer powe r Section V Replaceable Parts Sheet 3 of 4 Mfr J Stock...

Page 24: ...onnec tor male Can shield Boot cable Spacer standoff Lead probe ground includ e s R 5 Replaceable Parts Sheet 4 o f 4 Mfr 2 8480 2 8480 2 8480 i 99109 24655 24655 28480 28480 00866 28480 I F Stock No G 64B G 64C 8520 0017 456A 16A 8 120 0048 1 2 50 009 1 1 250 0092 4 56A 55A 4 12A 8 3A 0380 0048 4 56A 2 IA J rQ R S 1 0 1 0 1 0 1 1 1 0 1 0 1 0 1 0 1 0 1 0 1 1 See introduction to this section Model ...

Reviews: