![HP 10811A Operating & Service Manual Download Page 25](http://html.mh-extra.com/html/hp/10811a/10811a_operating-and-service-manual_161401025.webp)
Model 10811A/B
17
Table 4-1. Operational Verification Descriptions
1. Amplitude (0.55 volts (±0.05 volts/50
Ω
, 1 volt ± 0.2 volt/1k
Ω
).
a. Equipment
required:
BNC-to-BNC
Cable
500 and 1 K
Ω
loads
3406A AC Voltmeter
2. Harmonic Distortion (-25 dB below 10 MHz output signal).
a. Equipment
required:
141T Display Section
8553B Spectrum Analyzer - RF Section
8552B Spectrum Analyzer - IF Section
BNC-to-BNC
Cable
3. Spurious Signals - Signals unrelated to the 10 MHz output signal (-100 dB below the 10 MHz
output signal).
NOTE
Due to the complexity of measuring signal 100 dB
below a reference, this procedure will verify that no
spurious signals exist greater than 65 dB below the
normal 10 MHz output signal.
4.
a. Equipment
required:
141T Display Section
8553B Spectrum Analyzer - RF Section 8552B
Spectrum Analyzer - IF Section
BNC-to-BNC
Cable
5. Short Term Stability (5 X 10
-12
for 1-second averaging; 1 X 10
-11
for 10811-60105).
a. Equipment
required:
5065A Rubidium Vapor Frequency Standard
5061A Option 004, or 105A/B, See
HP Model 5390A may be used in place of the following:
10830A
Mixer
5345A Electronic Frequency Counter with Option 011
9825A Calculator with 98210A and 98213A or 98214A or 98216A
98034A HP-IB Interface
10515A Doubler
6-Plug AC Power Strip 3
BNC-to-BNC
Cables
6. Long Term Stability -<5 X 10
-10
over 24 hours; 1.5 X 10
-9
/24 hours for 10811-60105.
a. Equipment
required:
HP 1740A 100 MHz Oscilloscope,
105A/B Quartz Crystal Oscillator,
5065A Rubidium Frequency Standard, or 5061A Cesium Beam Frequency Standard.
.
4-11. In the following tests, the power supply voltages to the oscillator are not shown. They are
assumed to be:
Oscillator
Circuits ........................................ +12V dc
Oven Circuits ............................................... +20V dc
4-12. Any supply voltages within the specified limits listed in
may be used and will not affect
the operational verification. The voltages are only reference points and are used throughout this
manual.
Summary of Contents for 10811A
Page 2: ...Model 10811A B Page Intentionally Left Blank ...
Page 3: ...Model 10811A B ...
Page 9: ...Model 10811A B 1 Figure 1 1 10811A and 108118 Oscillators ...
Page 23: ...Model 10811A B 15 Page Intentionally Left Blank ...
Page 31: ...Model 10811A B 23 Page Intentionally Left Blank ...
Page 32: ...Model 10811A B 24 Table 4 4 Operational Verification Record ...
Page 44: ...Model 10811A B 36 Page Intentionally Left Blank ...
Page 46: ...Model 10811A B 38 Page Intentionally Left Blank ...
Page 48: ...Model 10811A B 40 Figure 8 1 Schematic Diagram Notes ...
Page 65: ...Model 10811A B 57 Table 8 2 Oven Controller Troubleshooting Tree ...
Page 70: ...Model 10811A B 62 Page Intentionally Left Blank ...
Page 71: ...Model 10811A B 63 ...
Page 72: ...Model 10811A B 64 Figure 8 15 Oven Controller Schematic Diagram ...
Page 73: ...Model 10811A B 65 ...