Hoyt 1652 IN Instruction Manual Download Page 14

Always check for voltage before testing and measuring on 

a circuit. This instrument is intended for measuring Low Ω and 

Insulation resistance on unenergized circuits only. 

Use the procedure explained at points 9.2, 9.3, 9.4.

Connect the test leads to the circuit to be measured and wait 

for a few seconds. The instrument will automatically discharge 

any remaining energy which could be present on the circuit, 

and will check for voltage at the same time.

Once you are sure that the circuit to be tested is not energized, 

then press the button for a short test duration or press and turn 

the button for a long test.

Once you end the test, allow a few seconds for the Instrument 

to automatically discharge the circuit.

- 12 -

9.7 Insulation Resistance Measurement @ 250Vdc

Summary of Contents for 1652 IN

Page 1: ...INSTRUCTION MANUAL ANALOGUE INSULATION TESTER 1652 IN...

Page 2: ...Features 4 Specifications 5 Why Test is Necessary 6 Instrument Layout 7 Simplified Instructions 8 Preparation for Measurement 9 Functions 10 Battery Fuse Replacement 11 Calibration Servicing 12 Clean...

Page 3: ...8 for details of built in warning features should your analogue insulation tester be connected to a live system 1 3 Never open your analogue insulation tester except for battery replacement see batter...

Page 4: ...of the voltage present will be heard DO NOT proceed to test and immediately disconnect the instrument from the circuit In addition the warning light will lit if the voltage is above 100Vdc or 70Vac W...

Page 5: ...ic discharge of capacitance and inductive circuit off charge stored in the circuit under test Live Warning and display of external voltage presence Fuse and overload protected On line battery monitori...

Page 6: ...0M Mid scale value 1M 2M 4M Scale Multiplier x1 2 x1 x2 Accuracy 5 Output short circuit current Approx 1 3mA Continuity Measuring range 3 500 Test leads fuse zero adjustment by knob Output short circu...

Page 7: ...ical conductors of electricity need to be insulated from each other so that they do not create electrical hazard or unnecessary consumption Badly insulated circuits can create leakage current which ca...

Page 8: ...6 TEST Button Continuity Zero OHM Adjust Function Selector HV Indication Battery OK ON Indication Testing terminals 6 Instrument Layout...

Page 9: ...finite towards zero If not the test leads or fuse 0 5A fast acting ceramic maybe faulty INSULATION TESTS M RANGES 1 Select the desired insulation test voltage range 250V 500V or 1000V 2 Connect the le...

Page 10: ...nsulation Mode may leave the circuits charged up if test leads are removed too quickly Avoid this by releasing the test button while the test leads are still connected to the circuit for a few seconds...

Page 11: ...9 2 DC Warning The DC warning buzzer will beep continuously when DC voltage is higher than 30Vdc on the test probes and the test button is Not pressed The neon light circuit live will lit when the vol...

Page 12: ...ment is intended for measuring Low and Insulation resistance on unenergized circuits only Use the procedure explained at points 9 2 9 3 9 4 The first procedure to follow is to zero the test leads and...

Page 13: ...to zero the test leads and the fuse resistance The instrument is equipped with a Zero Knob First short circuit the test leads by connecting them together then press the test button and adjust the zer...

Page 14: ...wait for a few seconds The instrument will automatically discharge any remaining energy which could be present on the circuit and will check for voltage at the same time Once you are sure that the cir...

Page 15: ...wait for a few seconds The instrument will automatically discharge any remaining energy which could be present on the circuit and will check for voltage at the same time Once you are sure that the cir...

Page 16: ...nd wait for a few seconds The instrument will automatically discharge any remaining energy which could be present on the circuit and will check for voltage at the same time Once you are sure that the...

Page 17: ...tteries i s low Disconnect the test leads from the Instrument remove the battery cover and the batteries Replace with six 1 5V R6 or L6 batteries taking care to observe the correct polarity Replace wi...

Page 18: ...ts If meter is not to be used for periods longer than 60 days remove the batteries and store them separately CAT IV Is for measurements performed at the source of the low voltage installation CAT III...

Reviews: