*   Electrical  connection:

  Dimension  of  all  wires  must  be  at  least  2,5  mm²  !

    

Brown

    =    ground                                                                                       (wire  no.  31)

    

       =   

  (for  charging  the  Stryker  PowerLoad 

Red 

permanent  positive

                         and  release  of  the  lateral  movement  device)

                         fused  in  the  vehicle  with  15A                                     

           (wire  no.  30)

    

  =   

,

                (wire  no.  15)                                    

Orange

ignition  positive

  fused  in  the  vehicle  with  30A   

     Never  connect  the  ignition  line  together  with  permanent  positive  !

     This  could  lead  to  consequential  damages,  which  are  not  covered  by  warranty  !

*     The  valve-control-circuit  is  internally  fused  with  5A

*    Max.  power  consumption  25A  at  12V  DC

HOVERBOARD

HOVERBOARD

Powerbase 

Technical  data

Disinfection  and  cleaning 

Phone   +43-660-  800  9000

Mail   

[email protected]

www.hover.at

*   Disinfection  and  cleaning  of  the  Hoverboard  shall  be  performed
     according  to  the  regulations  of  the  respective  country.

*   Never  use 

  or  other  corrosive  substances  !

disinfectants  containing  chloroacetic  acid

*   In  case  of  malfunctions,  please  contact  our  hotline

*    Custom  Hoverboard  for  Stryker  PowerPro  with  Stryker  PowerLoad

*    Height  lowered:                           135  mm  -  5.3"
      Height  when  active:                   220  mm  -  9.0"
      Height  for  reanimation  (rigid):   310  mm  -  12.2"

*    Standard  length:  2273  mm  -  89.5",  short  version:   2153  mm/84.8"

*    Total  weight  98/124  kg  -  216/273  lbs

*    Maximum  loading  capacity  520  kg  /  1146  lbs  (incl.  stretcher)

*    Ignition  AND  main  switch  ON:    Device  ready
      Ignition   OR   main  switch  OFF:  Device  lowered  (e.g.  for  loading)

Summary of Contents for Powerbase

Page 1: ...staff at all times The hoverboard should only be used for the purpose as described Please refer particularly to the instructions of the used stretcher The illustrations do not necessarily correspond t...

Page 2: ...lease option 4 Switch box contains a Relay 12V b Magnetic valves 12V c Terminal block 5 Air pressure tank approx 4 5 litres 1 2 gal with pressure switch 6 Reed switch 7 Blue main switch option 8 Red t...

Page 3: ...rbase Technical data Disinfection and cleaning Phone 43 660 800 9000 Mail info hover at www hover at Disinfection and cleaning of the Hoverboard shall be performed according to the regulations of the...

Page 4: ...st position e g for reanimation switch it on ATTENTION When lifting to the highest position the compressor runs continuously for approx 3 minutes Multiple lifting in short intervals will overheat the...

Page 5: ...0 1 0 1 0 2 5 2 5 1 0 1 0 2 5 2 5 1 0 1 0 2 5 2 5 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 1 0 2 5 1 0 2 5 1 0 1 0 1 0 1 0 1 0 1 0 1 0 Powerbase ground 31 bridge bridge bridge bridge bridge bri...

Page 6: ...h shock absorber 30621 without shock absorber 30311 air cushion 30057 magnetic valve block 30058 magnetic valve 1 30059 magnetic valve 2 30209 external compressor without Box 30291 level control valve...

Page 7: ...r 4 4 4 31021 quick connector 4 4 31022 quick connector 6 6 30520 relay 30530 relay holder 30072 Reed switch 30073 Block magnet 30091 Cable conduit 30500 connector 2 pins T f 30501 connector 3 pins T...

Page 8: ...tton 30244 pressure tube 4 mm red 30245 pressure tube 4 mm blue 30246 pressure tube 4 mm yellow 30240 pressure tube 6 mm red 30248 teflon tube white 31080 thermal protection tube red 30344 rubber grom...

Page 9: ...1041 toothcomb front 31011 LMD pneumatic cylinder 51062 locking bolt 31030 linear bearing 31040 knob Spare parts 30083 bellows without lateral movement device 30084 bellows with lateral movement devic...

Reviews: