SECTION 6 TROUBLESHOOTING
430-00587-008 (Rev. 2/05)
6 - 2
Plum XL Series
lists infusion system alarm codes that can be cleared by the operator. Also listed
are the alarm messages, descriptions, possible causes, and corrective actions.
Note:
Operational alarm messages are displayed on the LCD screen. Associated error
codes are displayed in the alarm history
(see Section 6.2.3.1, Alarm History)
Table 6-1.
Operational Alarm Messages and Corrective Actions
Error
Code
Alarm Message
Description
Possible Cause
Corrective
Action
01-1
OCCLUSION
Distal
occlusion
alarm
Distal pressure
above 10 psi
for five seconds
Unkink tubing,
check IV site,
or replace
administration
set
If condition
recurs,
01-2
Distal
occlusion
alarm
Distal pressure
above 10 psi
for two plunger
strokes
01-3
Distal
occlusion
alarm
Distal pressure
above 13 psi
01-4
Distal
occlusion
alarm
Excessive distal
pressure during
the valve leak test
03-1
Proximal
occlusion
alarm
Clamp closed;
tubing kinked;
possible occluded
tubing; defective
administration set;
or defective
pressure circuit
03-2
Proximal
occlusion
on secondary
during
backpriming
06-1
AIR IN LINE
BACKPRIMING
Air detected
in cassette
Proximal
air-in-line
1000 µL of air
has entered the
cassette since
last initialization
Backprime
to expel air
07-1
AIR IN LINE
Distal air-in-line
100 µL bolus
of air detected
at distal sensor
Remove and
reprime cassette
07-2
Distal air-in-line
260 µL of air
detected in
the last 2.6 mL
of fluid delivered
Remove and
reprime cassette