SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-00587-008 (Rev. 2/05)
7 - 30
Plum XL Series
Figure 7-14.
XL Series Pole Clamp
POLE CLAMP SHAFT/KNOB ASSEMBLY AND POLE CLAMP
SHAFT TIP REPLACEMENT
The recommended tool for this procedure is wide-head pliers.
The replacement parts for this procedure are:
Assembly, Shaft/Knob, Pole Clamp
Tip, Shaft, Pole Clamp
To replace the pole clamp shaft/knob assembly and the pole clamp shaft tip, refer to
, then proceed as follows:
1. Turn the control knob to OFF/CHARGE.
2. Disconnect the infusion system from AC (mains) power.
3. Turn the pole clamp shaft/knob assembly counterclockwise to remove it from the
pole clamp extrusion and loosen the pole clamp shaft tip from the pole clamp/shaft
knob assembly.
Note:
The pole clamp shaft tip has a long shaft that is pressed into the
threaded pole clamp shaft/knob assembly.
04K01037
POLE CLAMP
BACKING PLATE
ADHESIVE BACKED
INSULATOR
10-32 x 1/2
HEX HEAD SCREW
WITH WASHER (2)
POLE CLAMP
EXTRUSION
POLE CLAMP
SHAFT/KNOB
ASSEMBLY
POLE CLAMP
SHAFT TIP
REAR ENCLOSURE