SECTION 7 REPLACEABLE PARTS AND REPAIRS
430-00587-008 (Rev. 2/05)
7 - 28
Plum XL Series
Figure 7-13.
DataPort Assembly (International)
POLE CLAMP EXTRUSION, POLE CLAMP BACKING PLATE,
AND ADHESIVE-BACKED INSULATOR REPLACEMENT
Recommended tools and materials for this procedure are a medium size flat blade
screwdriver, a 5/16 inch nutdriver, and mild solvent.
The replacement parts for this procedure are:
Insulator, Adhesive-Backed
Screw, 10-32 x 1/2, Hex Head, Slotted
Plate, Backing, Pole Clamp
Extrusion, Pole Clamp
CAUTION:
Use proper ESD grounding techniques when handling components.
Wear an antistatic wrist strap and use an ESD-protected workstation. Store the PWA
in an antistatic bag before placing it on any surface.
To replace the pole clamp extrusion, pole clamp backing plate, and adhesive-backed
insulator, refer to
Figure 7-14, XL Series Pole Clamp
, then proceed as follows:
1. Separate the front enclosure assembly and rear enclosure assembly as described
2. Remove the power supply PWA as described in
DATAPORT
GASKET
REAR ENCLOSURE
DATAPORT
BRACKET
4-40 SELF-LOCKING
HEX NUT (2)
04K01039
DATAPORT/MCU
CABLE ASSEMBLY
4-40 x 5/16 LOCK
SCREW KIT (2)
DATAPORT/GROUND
CABLE ASSEMBLY
#4 SHOULDER
WASHER (2)
DATAPORT
CAP
4-40 x 1/4
HEX HEAD
SCREW (2)