![horiba OM-71G Instruction Manual Download Page 38](http://html1.mh-extra.com/html/horiba/om-71g/om-71g_instruction-manual_2162274038.webp)
Dissolved oxygen measurement
18
• Setting air pressure correction (default: 1013 hPa)
Amount of oxygen that can be dissolved in water is affected by air pressure in the loca-
tion where measurement is taking place. You need to set the air pressure of the mea-
surement location to the instrument before calibration to correct the pressure influence,
and then perform measurement.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "PRES" (air pressure collection setting) and
then press the
key.
3. Enter the air pressure value of the measurement location and then press the
key.
4. Press the
key to reflect the change.
key
key
Summary of Contents for OM-71G
Page 2: ......
Page 20: ...M E M O...
Page 64: ...M E M O...
Page 74: ...M E M O...
Page 82: ...M E M O...
Page 84: ......
Page 85: ......
Page 86: ......