background image

LAQUAtwin series (Na

+

) Operation

5

CODE: GZ0000460286

Setup Mode

The setup mode allows the user to customize the meter to his specific needs.
To enter the setup mode, press and hold the MEAS and ON/OFF switches for over 3 seconds when the
meter is switched OFF. All the LCD segments appear and then the meter enters the setup mode.

Tip

To have the changes apply, you need to go through the entire steps from “Setup mode entry” to
“Setup completion” shown below. To leave a setting as it is, just press CAL switch in the setting.

To exit the setup mode with no change of settings, press the ON/OFF switch earlier than pressing
CAL switch in the last step but one, or the “Backlight setting” step.

Setup mode entry 

Unit setting

The display units can be changed.

1st calibration point setting

The concentration of the 1st calibration point can be set.

Note

If you changed either of the calibration
concentration settings, calibrate again at
the both set concentrations before mea-
surement. When either calibration concen-
tration is changed, the calibration data are
initialized.

2nd calibration point setting

The concentration of the 2nd calibration point can be set.

* In calibration point setting, pressing the 

MEAS switch increases the displayed 
value. After the displayed value reaches 
9800, it returns to 3.

All segment

indication (2 s)

Software ver.

indication (1 s)

ON/OFF switch
MEAS switch
CAL switch

ppm (default)

mg/L

1st cal. point

*

2nd cal. point

*

Measurement mode change

The measurement mode can be switched.

*  Measurement display change is available in 

the AS mode. Refer to " 

 Measurement 

display change" (page 3).

Multiplying compensation setting

The coefficient (0.01 to 9.90) to be applied to the measured value can be set. The compensated result is
displayed as the measured value. The default setting is 1.00.

Backlight setting

The backlight can be switched to ON or OFF.

Setup completion

AS mode* (default) 

AH mode

Backlight ON 

(default)

Backlight OFF

END indication (1 s)

Powered OFF

For any questions regarding this product, please contact 
your local agency, or inquire from the following website. 
http://global.horiba.com/contact_e/index.htm

 

31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto, 
601-8306, Japan
http://www.horiba-adt.jp

Summary of Contents for LAQUAtwin-Na-11

Page 1: ... The closeness of agreement between a mea sured value and an actual value of standard solu tion after two point calibration using 150 ppm and 2000 ppm standard solutions The standard solution used for the later cali bration was measured after two point calibra tion at 150 ppm and 2000 ppm The calibration and measurement are per formed at the same temperature The error of standard solutions and rou...

Page 2: ...asurement mode automatically 5 Open the light shield cover and remove the standard solution Then remove moisture on the sensor by gently dabbing with a soft tissue This completes the 1st point calibration 6 To perform 2nd point calibration repeat steps 2 to 5 Calibration error If blinks and Er4 error dis play appears the calibration has failed Perform electrode conditioning Check that the correct ...

Page 3: ...perature sensor adjustment To perform accurate measurement with correc tion for temperature effects follow the steps below Normally this is not necessary 1 Ready a reference thermometer and allow the meter and reference thermometer to reach to room temperature Sampling sheet holder cover Part No 3200459736 Light shield cover Sampling sheet B Part No 3200053858 2 Set the display mode to temperature...

Page 4: ...same time Er4 is dis played during the calibra tion Please note that if you press the CAL switch in mV or temperature display mode Er4 is displayed This is because there is no calibration facility available for these modes Er1 is dis played soon power ON The internal IC in the meter may be defective Perform meter initializa tion If Er1 is still displayed after the initial ization the internal IC i...

Page 5: ... concen tration is changed the calibration data are initialized 2nd calibration point setting The concentration of the 2nd calibration point can be set In calibration point setting pressing the MEAS switch increases the displayed value After the displayed value reaches 9800 it returns to 3 All segment indication 2 s Software ver indication 1 s ON OFF switch MEAS switch CAL switch ppm default mg L ...

Reviews: