Basic Operation

Calibration

Calibration is required before measurement.
Use standard solution within the measurement
range in the specifications. See also the tip of
“How to select standard solution” below.

Tip

Calibration values are saved even if the meter
is switched OFF.

Calibration value is rewritten if calibration is
repeated using the same standard solution.

Tip

How to select standard solution:
The following table show an example of the rela-
tionship between the concentrations of targeted
samples and standard solution used for calibra-
tion.

For accurate measurement, calibrate using 2
points of standard solution with a tenfold or more
concentration difference encompassing the tar-
get concentration.
When the concentration to be measured is very
high or very low, accuracy may be poorer.

Measurement

Power ON

1. Press and hold the ON/

OFF switch.

The power is switched ON,
and the meter model
number is displayed on the
LCD.

Power OFF

1. Press and hold the ON/OFF switch.

The power is switched OFF.

Calibration points

The number of calibration points is up to 2.

Multi-point calibration

1. Set the concentrations of standard solution

for calibration referring to " 

 1st calibra-

tion point setting" (page 5) and " 

 2nd cal-

ibration point setting" (page 5).

The 1st point is set to 150 ppm and the 2nd
point is set to 2000 ppm by the default.

2. Open the light shield cover and place some

drops of the standard solution on the flat
sensor taking care to cover the entire flat
sensor. 

Rinsing the sensor with the standard solution
beforehand will provide a more accurate
calibration as it will reduce sample crossover
contamination.

3. Close the light shield cover and press the

CAL switch.

The meter enters the CAL mode and blinks the
display of the set 1st-point concentration.
Pressing the MEAS switch switches the
displayed value between the set
concentrations.

4. With the set concentration of the 1st point

displayed, press the CAL switch. 

 and 

 blink, and the calibration value is

displayed.
After the calibration is complete, 

 and 

stop blinking and the measured value is
displayed. 
The calibration value at 25

C is displayed for 1

s and the display returns to the measurement
mode automatically.

5. Open the light shield cover and remove the

standard solution. Then remove moisture
on the sensor by gently dabbing with a soft
tissue.

This completes the 1st point calibration.

6. To perform 2nd point calibration, repeat

steps 2. to 5.

Calibration error

If 

 blinks and Er4 (error dis-

play) appears, the calibration
has failed. 
Perform electrode conditioning. 
Check that the correct standard solution is used,
and repeat calibration after cleaning the sensor.
If the calibration repeatedly fails when using the
correct standard solution(s), the sensor may
have deteriorated. Replace the sensor with new
one.

Light shield cover

[unit: ppm]

Targeted sample 

concentration

Standard solution

Low

High

2000

2000

9000

150

150

2000

15

 15*

150

 *How to prepare 15 ppm standard solution:
Mix one part 150 ppm standard solution to 9 parts 
purified water (1:9 volume ratio).

Sample setting

1. Open the light shield cover and put some

drops of sample on the flat sensor to cover
the entire flat sensor.

2. Close the light shield cover.

Measurement mode

The auto stable (AS) mode and the auto hold
(AH) mode can be selected. Refer to " 

 Mea-

surement mode change" (page 5) for the opera-
tion to set the measurement mode.

Auto stable (AS) mode

This is the default setting. 

 appears when

the measured value meets the stability criteria.
If the value changes, 

 disappears.

1. Confirm that the meter is in the measure-

ment mode, and place a sample on the sen-
sor.

When the read value meets
the stability criteria, 
appears and the reading is
locked.

2. Document the displayed value when 

appears.

If the read value does not meet the stability
criteria, 

 disappears and the reading

changes with time.

Auto hold (AH) mode

 appears when the measured value meets

the stability criteria. The reading then locks
and will not change until the MEAS switch is
pressed for the next measurement.

1. Confirm that the meter is in the measure-

ment mode, and place a sample on the sen-
sor.

2. Press the MEAS switch.

The auto hold function is
activated. 

 blinks until the

measured value has
stabilized.

When the measured value is stable, 

 stops

blinking and the displayed value is locked with

 and 

 displayed simultaneously.

3. Document the displayed value.
4. 
Press the MEAS switch.

The auto hold function is deactivated and 
disappears.
Be sure to perform this step before starting the
next measurement. Or, you may mistake the
displayed hold value for the next measured
value.

Summary of Contents for LAQUAtwin-Na-11

Page 1: ... The closeness of agreement between a mea sured value and an actual value of standard solu tion after two point calibration using 150 ppm and 2000 ppm standard solutions The standard solution used for the later cali bration was measured after two point calibra tion at 150 ppm and 2000 ppm The calibration and measurement are per formed at the same temperature The error of standard solutions and rou...

Page 2: ...asurement mode automatically 5 Open the light shield cover and remove the standard solution Then remove moisture on the sensor by gently dabbing with a soft tissue This completes the 1st point calibration 6 To perform 2nd point calibration repeat steps 2 to 5 Calibration error If blinks and Er4 error dis play appears the calibration has failed Perform electrode conditioning Check that the correct ...

Page 3: ...perature sensor adjustment To perform accurate measurement with correc tion for temperature effects follow the steps below Normally this is not necessary 1 Ready a reference thermometer and allow the meter and reference thermometer to reach to room temperature Sampling sheet holder cover Part No 3200459736 Light shield cover Sampling sheet B Part No 3200053858 2 Set the display mode to temperature...

Page 4: ...same time Er4 is dis played during the calibra tion Please note that if you press the CAL switch in mV or temperature display mode Er4 is displayed This is because there is no calibration facility available for these modes Er1 is dis played soon power ON The internal IC in the meter may be defective Perform meter initializa tion If Er1 is still displayed after the initial ization the internal IC i...

Page 5: ... concen tration is changed the calibration data are initialized 2nd calibration point setting The concentration of the 2nd calibration point can be set In calibration point setting pressing the MEAS switch increases the displayed value After the displayed value reaches 9800 it returns to 3 All segment indication 2 s Software ver indication 1 s ON OFF switch MEAS switch CAL switch ppm default mg L ...

Reviews: