Conductivity measurement
18
●
Setting the instrument
• Setting the temperature display (default: ATC)
When a conductivity cell with a temperature sensor is used, or a conductivity cell without
a temperature sensor is used with a temperature electrode, the automatic temperature
measurement function can be used. During measurement, the temperature sensor
measures the temperature of the sample and displays the result on the instrument. If
automatic temperature measurement function is not used, or the temperature connector
is not connected to the instrument, the temperature set in the instrument is displayed.
1. Press the
key to enter the setting mode.
2. Select the "TEMP" (temperature setting) and then press the
key.
3. Press the
keys to select the "ATC" (automatic temperature measurement)
or the "MTC" (manual temperature display) and then press the
key.
4. If you select the "MTC", enter the temperature to be displayed and then press
the
key.
5. Press the
key to reflect the change.
key
< When MTC is set >
key