Chapter 3 Basic Operation
3.5 Data operation
60
3.5.6 pH check history display
The pH check history can be printed out.
●
Repeatability check of calibration
After calibration is completed, immerse the electrode in
the standard solution of pH7 and press the ENTER key
to perform the repeatability check of calibration.
It is necessary to calibrate the standard solution of pH7
in NIST or USA of the standard solution setting. This
operation is valid only once after calibration.
●
Calibration repeatability check
1.
Press the DATA key to enter the data operation
screen.
2.
Press the
▲
or
▼
key to enter the pH check history
display screen.
3.
Press the ENTER key.
The repeatability of the latest check data is
displayed.