Set Up
6
Fixing the Rear Cover
There are small protrusions in the rear cover and the main unit case. To tighten the cover,
align these protrusions and screw straight. If they are obliquely inserted, the cover cannot be
colsed completely.
Mounting the Hood
To mount the hood (FD-10) provided as an option, use screws on the opposite side where
metal fitting for 50 A pipe is attached.
Air Purge
The two tapped holes of PF 1/8 screw are the side of STD/SLOPE control on the bottom of the
transmitter. The tapped holes are used to be the inlet and outlet ports of air purge after boring
a hole to the inside of the transmitter and connecting the air pipe joints. The air purge is
required in the environment of corrosive gas or huge humidity.
Hood