![horiba HEMAPREP RAB316AEN User Manual Download Page 37](http://html1.mh-extra.com/html/horiba/hemaprep-rab316aen/hemaprep-rab316aen_user-manual_2162217037.webp)
3.7.
Incomplete Pickup of Blood Drop
Cause
Solution
Spreader blade is not in contact with blood drop.
The drop was delivered behind the target area.
Deliver the drop of blood directly on the target location.
If insufficient blood is still picked up, deliver the drop slightly
before the target location (in direction of the spreader blade).
3.8.
Drop of Blood Does not Spread Out
The drop of blood does not spread out when delivered and moisten the slide.
Cause
Solution
Slide is dirty.
Replace the slide.
Blood dispenser is held at an angle which is too
vertical.
Hold dispenser at about a 45° angle when the blood drop is
dispensed. The amount of blood which is delivered can be
controlled by the angle at which the dispenser is held.
Maintenance and Troubleshooting
Troubleshooting Procedures
User Manual
Ref: RAB316AEN
33
Summary of Contents for HEMAPREP RAB316AEN
Page 1: ...Blood Spreading System User Manual Ref RAB316AEN...
Page 2: ......
Page 3: ...User Manual User Manual Ref RAB316AEN...
Page 4: ......
Page 6: ...ii User Manual Ref RAB316AEN...
Page 12: ...Foreword Legal Information 8 User Manual Ref RAB316AEN...
Page 30: ...Workflow Performing the Test 26 User Manual Ref RAB316AEN...
Page 38: ...Maintenance and Troubleshooting Troubleshooting Procedures 34 User Manual Ref RAB316AEN...
Page 40: ...36 User Manual Ref RAB316AEN...