Resetting to factory default settings
48
HORIBA
●
Resetting to factory default settings
The instrument settings can be reset to the factory default settings. The calibration data
is deleted but the data of date and time, and the saved data are not deleted. Make sure
there will be no problems before using this function. When this function is used, the
temperature calibration data is also initialized.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
3. Press the
keys to select the "INIT" (initialize) and then press the
key.
4. Select "YES" to initialize the settings to the factory default settings, or select
"NO" to cancel initialization. And then press the
key.
When "YES" is selected, "END" appears after the settings are initialized. Press the
key.
To return to the setting mode, press the
key.
key
key
key
key
Selected "YES"
and press the
key