![horiba D-72 Instruction Manual Download Page 32](http://html1.mh-extra.com/html/horiba/d-72/d-72_instruction-manual_2162200032.webp)
pH measurement
16
HORIBA
3. Press the
keys to select the "ATC" (automatic temperature compensation)
or the "MTC" (manual temperature compensation) and then press the
key.
4. If you select the "MTC", enter the temperature to be compensated for and then
press the
key.
• Setting the standard solution used for calibration (default: NIST)
Set the standard solution used for calibration. With this instrument, you can choose from
3 types, NIST, USA, and CUST (the standard solution other than NIST and USA).
Set according to the standard solution to be used.
1. Press the
key to enter the setting mode.
2. Select the "BUFFER" (pH standard solution setting) and then press the
key.
3. Press the
keys to select "NIST", "USA", or "CUST" according to the
standard solution to be used and then press the
key.
< When MTC is set >
key
key
key
key
Summary of Contents for D-72
Page 1: ...Portable pH ORP Meter D 72 Instruction Manual CODE GZ0000333604...
Page 2: ......
Page 16: ......
Page 78: ...M E M O 62 HORIBA...
Page 86: ...M E M O 70 HORIBA...
Page 87: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...
Page 88: ......