5 Reference
5.2 Specifications
76
HORIBA
5.2
Specifications
●
Measurement target
●
Items in common among meter models
Target
Item
Description
pH
Measurement
principle
Glass electrode
Display range
pH -2.00 – 16.00
Measurement
range
pH 0.00 – 14.00
Resolution
0.01 pH
Repeatability
±0.01 pH ±1digit
Temp.
Measurement
principle
Thermistor
Measurement
range
0.0 – 100.0 ºC
Resolution
0.1 ºC
Repeatability
±0.1 ºC ±1digit
Data memory
capacity
Max. 300 pieces of data
Power
Dry cell batteries type:AA alkaline
with automatic power OFF function
Ambient
temperature
0 – 45 ºC
Dimensions
170(H)
×
80(W)
×
40(D) mm
Mass of main unit
(including
batteries)
300 g