4 Maintenance and Troubleshooting
4.2 Troubleshooting
60
HORIBA
●
ERR No.06 Max. calibration points exceeded
Explanation
Calibration was performed on a fourth calibration point.
This error does not affect calibration data obtained from
previous three calibration points.
Refer to “
Standard solution calibration” page 21.
Cause
How to solve problem
Calibration was
performed on a
fourth calibration
point.
Limit the number of calibration
points to three.
This error message is cleared by
setting the pH meter to
Measurement mode.