Maintenance &
Troubleshooting
User Manual - RAB237AEN
7–41
6.2. Logs
◆
When the instrument detects an error, the status button turns orange and the error message is
displayed in the "
Logs
" tab.
◆
To access the "
Logs
" tab:
1-
Press "
Maintenance menu
" button (
).
2-
Press "
Logs
" tab.
◆
The log messages are sorted by category:
- See
.
.
6.2.4. Maintenance logs, page 7-47
6.2.5. Startup logs, page 7-47
6.2.1. Alarms logs
Diag.7-47 Maintenance menu button
Message
Possible cause/Corrective action
Startup cycle failed
The cycle has not been able to be completely performed or no
result received or no previous result to display:
- Run a startup cycle by pressing the "
Startup
" button.
Invalid startup cycle
The startup cycle has correctly carried out, but the result of
the blank cycle is out of acceptable limits or the delay
between 2 analyses (X hours adjustable) is exceeded or the
technician asks for a restart without Shutdown:
- Run a startup cycle by pressing the "
Startup
" button.
Printing blank report failed
Print error or report generation error:
- Check if printer is on line. If not, switch it on.
- Check if there is enough paper. If not, See
ticket printer paper changing, page 7-16
.
- Check printer settings (See Section 6,
Analyser communication failed
A communication error with the analyser occurred:
- Run a startup cycle by pressing the "
Startup
" button.
Analyser connection failed
Connection not initialized between analyser and application
(analyser do not answer or analyser version is not compatible
with the application version):
- Restart the instrument.
- Check version numbers (See
).
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Critical cycle error
- Run a startup cycle by pressing the "
Startup
" button.
Application error
A software problem occurs during data saving or file opening
for example:
- Restart the instrument.
Reagent temperature error
A reagent is not at the operating temperature:
- Run a startup cycle by pressing the "
Startup
" button.
- Replace the reagent or the pack.
See
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...