6.2.3. QC logs
6.2.4. Maintenance logs
6.2.5. Startup logs
Door opening mechanic cycle failed
Error during door opening check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Needle positioning mechanic cycle
failed
Error during needle motion check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Tube piercing mechanic cycle failed
Error during tube piercing motion check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Valve positioning mechanic cycle failed
Error during valves check cycle
- Restart the instrument.
Waste piston positioning mechanic
cycle failed
Error during waste piston motion check cycle
- Manually check the mechanical movement
- Run a startup cycle by pressing the "
Startup
" button.
Cleaning service cycle failed
- Check cleaner level.
- Restart the instrument.
- Run a startup cycle by pressing the "
Startup
" button.
Concentrated cleaning service cycle
failed
- Check cleaner level.
- Restart the instrument.
- Run a startup cycle by pressing the "
Startup
" button.
Draining service cycle failed
- Restart the instrument.
- Run a startup cycle by pressing the "
Startup
" button.
All reagents prime service cycle failed
- View the chambers for reagent priming.
- Verify that the reagent is coming in and draining out of the
chambers when a cycle is initiated.
- Verify that there are no air bubbles in the reagent lines when
priming.
Message
Possible cause/Corrective action
Cannot open control configuration
file, the application must be
reinstalled, contact your local
technical support.
QC list recovery impossible.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Message
Possible cause/Corrective action
Cannot save closed reagent, please
contact your local technical support.
Internal read/write error:
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Software internal error, please contact
your local technical support.
- Contact your local HORIBA ABX Technical Support
Representative for assistance.
Message
Possible cause/Corrective action
Invalid startup cycle (wbc/rbc/hgb/
plt)
Parameters WBC , RBC , HGB & PLT > startup limits
- Run a startup cycle by pressing the "
Startup
" button.
Message
Possible cause/Corrective action
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...