pH Electrode

9615-10D Instruction Manual

Thank you for purchasing a HORIBA pH electrode.
Read this manual carefully, before using the electrode.

J

Caution on Use

 

Safety Precautions

 

Points of concern

z

Do not shock the electrode.

z

Do not measure chapped or cracked electrode. Replace
such electrode with a new one.

z

Do not allow the connector to come in contact with water
or unclean hands.  If not, accurate measurement cannot
be performed.

z

Always use the 3.33 mol/L-KCl solution (model 300)
internal solution for reference electrode.

z

If the responsive glass membrane or liquid junction is
very dirty and cannot be washed clean using pure water,
cleanse it depending on dirt condition. (Refer to “
Maintenance ” (page 3).)

z

Avoid continuous use of this electrode at acidic or
alkaline concentrations above 0.1 mol/L. Using the
electrode under these conditions may harm
performance and shorten the life of the electrode. 

z

Do not measure the solutions which contain fluoride ion,
which may corrode the glass electrode.

z

Do not suddenly immerse the electrode in samples
where the temperature variation is 50

°

C or more.

z

Silver chloride can be detected inside the responsive
glass membrane. However, it will not cause any
performance problem.

z

Do not use the electrode in any place where usable
temperature is out of the specified range.

z

White crystal can adhere to the internal solution filler
port or protective cap after storage of the electrode.
However, it will not cause any performance problem. 

J

Packaged Contents

CAUTION

Chemical solution

The internal solution in the electrode uses potassium chloride 
(3.33 mol/L-KCI). If the internal material comes in contact with the 
hands or skin, wash immediately with water. If the internal solution 
comes in contact with the eyes, flush with a large amount of 
running water, then seek medical advice.

Glass fragments

Glass fragments can cause injury.
The outer tube and tip of the electrode are made from glass.
Be careful not to break them.

Name

Q’ty

Electrode

 

9615-10D

1 pc

Attachment holder

1 pc

Instruction manual (this book) 1 booklet

J

Specifications and Parts Description

 

Part name

 

Specifications

J

 Connecting to pH meter

1. Insert the electrode connector into the connector port

sleeve on the meter, after aligning with the pin. Do not
insert the connector unless it is aligned properly with the
connector port.

2. Press the electrode connector into the connector port on

the meter, while turning the connector to the right.

3. Slide the connector cover over the connector. Then,

push the cover in straight until it comes in light contact
with the meter case. Do not turn the cover.

4. Insert the temperature connector into the jack on the

meter. Insert the connector firmly, until the O-ring on the
connector can no longer be seen.

Note

Be sure to install the temperature connector. If not, accurate
measurement cannot be performed.

Electrode model

9615-10D

Measurement range

pH 0 to 14

Usable temperature range

0

°

C to 100

°

C

Storage temperature range

0

°

C to 50

°

C

Internal electrode

Silver/silver chloride

Reference electrode internal 
solution

3.33 mol/L-KCl solution (model 300)

Liquid junction material

Porous ceramic

Wetted material

Glass, ceramic

Electrode length

198 mm

Ext. diam. of wetted part

12 mm

Wetted part height

Approximately 13 mm (from the tip of 
electrode)

Lead length

1 m

Liquid
junction

Protective cap

Stem

Cap

Temp. connector

Electrode connector

Connector cover

Lead

Responsive glass
membrane

Internal solution 
filler port

Knurled nut