Conductivity electrode (9383-10D)

Instruction Manual

Thank you for purchasing the HORIBA electrode. Read this
manual carefully before using the electrode.

Conformable standards

Authorised representative in EU

HORIBA UK Limited
Kyoto Close Moulton Park Northampton NN3 6FL UK

Cautions during handling

Do not allow the electrode to come in contact with any
hard surface.

If the electrode pole plate is dry, immerse the electrode
in pure (ion exchange) water for at least one hour, prior
to use.

The electrode connector requires high insulation. Do not
allow the connector to be in contact with water or dirty
hands.

Do not rub the electrode plate with a brush or polish it
with a polishing agent.

The water-resistant construction of meters can be used
in combination with this electrode to provide water-
resistant construction (conforming to IP-67). When
measuring, do not immerse the cap, the lead or the
connector in the sample.

Packaged contents

Specifications and names of parts

Names of parts

Specifications

This equipment conforms to the following standards:

RoHS:

EN50581
9. Monitoring and control instruments

Name

Q’ty

Electrode

1 pc

Operation Manual

1 copy

Model

9382-10D

Cell constant

100 m

-1

 (Former units: 1 cm

-1

)

Measurement range

0.1 mS/m to 10 S/m (Former units: 1 

μ

S/cm to 

100 mS/cm)

Usable temp. range

0

°

C to 80

°

C

Storage temp. range

0

°

C to 50

°

C

Electrode material

Titanium - platinum black

Wetted part materials

PPS, PSF and titanium

Electrode 
connector

Connector cover

Cap

Lead

Temp. connector (T electrode)

Protective cap

Electrode plate

Preparations

Connecting to meter

1. Insert the electrode connector into the connector port

sleeve on the meter, after aligning with the pin. Do not
insert the connector unless it is aligned properly with the
connector port.

2. Press the electrode connector into the connector port on

the meter, while turning the connector to the right.

3. Slide the connector cover over the connector. Then,

push the cover in straight until it comes in light contact
with the meter case.
Do not turn the cover.

4. Insert the temperature connector into the jack on the

meter. Insert the connector firmly, until the O-ring on the
connector can no longer be seen.

Setting cell constant

The cell constant for this electrode is displayed on the
electrode cap.
Enter the cell constant (shown as 0.954 x 100, in this
example) of the electrode, by following the instructions in the
Operation Manual for the meter.

Example: 

9383-10D

 

LOT

. ......

0.954 x 100 m

-1

Preparing electrode

Either wash the electrode plate using a bottle that contains
pure (ion exchange) water, or immerse the electrode in a
beaker containing pure (ion exchange) water and lift the
electrode up and down a few times to rinse it, then wipe it
dry using filter or tissue paper.

Electrode length

150 mm (incl. cap)

Max. height of electrode 
plate

53 mm (position from electrode tip)

Ext. diam. of wetted part 

16 mm

Lead length

1 m

C1

T1

T2

C2

Summary of Contents for 9383-10D

Page 1: ...7 水温と機器の温度差が 5 C 以内で水深 1m 30 分水没しても内部に水が侵入しない 内容物 仕様と各部の名称 各部の名称 仕様 名称 数量 電極 1 本 取扱説明書 1 部 形式名 9383 10D セル定数 100 m 1 旧単位 1 cm 1 測定範囲 0 1 mS m 10S m 旧単位 1μS cm 100 mS cm 使用温度範囲 0 80 C 保存温度範囲 0 50 C 極材 チタン 白金黒 接液部材質 PPS PSF チタン 電極長さ 150 mm キャップを含む 極板最上部高さ 53 mm 電極先端からの位置 接液部外径 16 mm リード長 1 m 電極コネクタ コネクタカバー キャップ リード線 温度コネクタ T 極 保護キャップ 極板 準備 計器への接続 1 計器本体のコネクタ受け口のピンに合わせて電極コネ クタの溝を差し込みます 溝に合わないうちに無理に押 ...

Page 2: ...い 高濃度サンプルを測定後に低濃度サンプルを測定する場合 特 に電極をよく洗い 高濃度サンプルの影響がないことを確認し てからサンプルに浸けてください 保守 電極に付着したサンプルを純水でよく洗い落としてく ださい 電極の汚れがひどく純水で洗っても落ちない場合 状況 に応じて下記の洗浄を行い その後純水で十分にすすぎ 洗いをしてください 長期間の使用において 極板の表面状態が変化すること によりセル定数が変動することがあります 2 3 か月 に一度を目安にセル定数の測定をお勧めします 詳しく は 計器の取扱説明書を参照ください 一般の汚れ 油分の汚れ 中性洗剤に浸し すすぎ洗いしてください 無機成分などの汚れ 5 程度の硝酸中に約 30 分浸けてください 極板最上部 サンプル 保管 1 電極に付着したサンプルを純水で良く洗い落とします 2 保護キャップの内部を純水で洗い 極板最上部が浸る程...

Page 3: ...Former units 1 cm 1 Measurement range 0 1 mS m to 10 S m Former units 1 μS cm to 100 mS cm Usable temp range 0 C to 80 C Storage temp range 0 C to 50 C Electrode material Titanium platinum black Wetted part materials PPS PSF and titanium Electrode connector Connector cover Cap Lead Temp connector T electrode Protective cap Electrode plate Preparations Connecting to meter 1 Insert the electrode con...

Page 4: ...refully using pure ion exchange water to remove any sample still clinging to the electrode If the electrode is very dirty and cannot be washed correctly using pure ion exchange water wash it using the appropriate method below Then rinse the electrode well using pure ion exchange water A long term use of the electrode may result in shifts of the cell constant due to changes in the surface condition...

Reviews: